- Topic ID: id_11038923
- Version: 5.0
- Date: Jan 20, 2020 8:36:42 PM
Save Restore Generator Runtime Parameters
Prerequisites
Overview
This document defines the procedures to save and restore the Generator Runtime Parameter values, which are used during Generator failure identification and troubleshooting.
1 Save Generator Runtime Parameters
Procedure
- Select DIAGNOSTICS from Service Desktop.
- Select GENERATOR TOOL – JEDI, or X-RAY GENERATION, GENERATOR TOOL – JEDI.
- Click SAVE RUNTIME PARAMETERS.
- Click OK in Attention window.
- Click OK in Save Runtime Parameters window.
- Click DISMISS in Generator Tool - JEDI window.
- Save System State.
2 Restore Generator Runtime Parameters
Procedure
- Select DIAGNOSTICS from Service Desktop.
- Select GENERATOR TOOL – JEDI, or X-RAY GENERATION, GENERATOR TOOL – JEDI.
- Select RESTORE RUNTIME PARAMETERS.
- Click OK in Attention window.
- Click OK in Restore Runtime Parameters window.
- Click DISMISS in Generator Tool - JEDI window.