• Topic ID: id_17423162
  • Version: 2.0
  • Date: Mar 6, 2020 9:59:11 PM

RAID Performance Measurement

Prerequisites

Overview

This procedure describes how to measure the scan data disks RAID performance by using io_raid_perf command.

The io_raid_perf command is available on Optima CT680 with software version 16BW45.x or later.

Procedure

  1. Shutdown application software
  2. Open a Unix shell and type the following:

    {ctuser@tc10bay} io_raid_perf

  3. The following outputs are displayed on console continuously (every 320MB packets, 0 to 2880MB as one set.).

    {ctuser@tc10bay} io_raid_perf


                    0,1409789210.802099,  0.000000, 10,32.000,0.540,0.540,0.000,592.154,45.49%,65.57%
    320,1409789211.464679,  0.662580, 10,32.000,0.662,0.662,0.000,483.063,45.56%,65.57%
    640,1409789212.136187,  1.334088, 10,32.000,0.671,0.671,0.000,476.613,45.63%,65.57%
    960,1409789212.827031,  2.024932, 10,32.000,0.691,0.685,0.006,463.267,45.69%,65.57%
    1280,1409789213.424901,  2.622802, 10,32.000,0.598,0.598,0.000,535.320,45.76%,65.57%
    1600,1409789214.107648,  3.305549, 10,32.000,0.683,0.683,0.000,468.765,45.82%,65.57%
    1920,1409789214.685581,  3.883482, 10,32.000,0.578,0.578,0.000,553.779,45.89%,65.57%
    2240,1409789215.340814,  4.538715, 10,32.000,0.655,0.655,0.000,488.449,45.95%,65.57%
    2560,1409789215.936413,  5.134314, 10,32.000,0.596,0.595,0.000,537.361,46.02%,65.57%
    2880,1409789216.615927,  5.813828, 10,32.000,0.679,0.679,0.000,470.996,46.08%,65.57%
                  

    Each data element can be translated as follows.

  4. The 9th data element shows RAID performance in MB/sec. Calculate the average value of 5 data at least. In case of Table 1, the average of the first five data packets is 510.083 MB/sec. The specification is 360MB/sec or higher.

Finalization