Open Console (Z840) RAID Controller Card Replacement
Prerequisites
Overview
This procedure shall be followed when replacing the RAID Controller Card in the host computer (Z840).
1 Preparation
Procedure
- Shutdown system. Select one of the following methods to Power
OFF the Console:
-
If Applications are up, click on the [Shut Down] button on desktop display and select [Shutdown].
-
If Applications are down, open a Terminal Window. Type: halt , then press ENTER.
-
When halt command has finished, power Off the console at the front panel switch.
-
- Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.
2 RAID Controller Card Replacement
Procedure
- Remove the side access panel by loosen three screws and pull
out the latch to release the computer left side cover.
Figure 1. Side Access Panel Removal

- Remove the following components from the host computer (Z840):
-
Expansion Card Support
-
Airflow Guide
Figure 2. Z840 Airflow Guide and Expansion Card Support

-
- Locate RAID Controller Card FRU by referencing the Illustration
below.
Figure 3. Z840 Component Location

- Disconnect the two SAS cables from the RAID Controller Card. Record the port location.
- Replace the existing RAID Controller Card (Slot 4) with new
one. Pull-up the latch when removing the card.
Figure 4. RAID Card Latch

- Reconnect the SAS cables onto the RAID Controller Card.
- Reinstall all removed components, and reconnect any cables that have been disconnected.
3 Restore the Console
Procedure
- Remove LOTO on console.
- Reinstall the host computer side cover.
4 Finalization
Procedure
- Power on console and execute “Recreate Scan Disk Array”
from Reconfig menu.note:
All scan date will be lost.
- Perform the Functional Checks → System Scanning Test instructions from the procedure list.