• Topic ID: id_17423188
  • Version: 4.0
  • Date: Aug 10, 2021 9:56:50 PM

Open Console (Z840) Power Supply Replacement

Prerequisites

Overview

This procedure shall be followed when replacing the Power Supply in the Z840 computer.

1 Preparation

Procedure

  1. Shutdown system. Select one of the following methods to Power OFF the Console:
    • If Applications are up, click on the [Shut Down] button on desktop display and select [Shutdown].

    • If Applications are down, open a Terminal Window. Type: halt , then press ENTER.

    • When halt command has finished, power Off the console at the front panel switch.

  2. Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.

2 Power Supply Replacement

Procedure

  1. Remove the side access panel by loosen three screws and pull out the latch to release the computer left side cover.

    Figure 1. Side Access Panel Removal

  2. Remove the following components from the host computer (Z840):
    • Expansion Card Support

    • Airflow Guide

      Figure 2. Z840 Airflow Guide and Expansion Card Support

  3. Locate Power Supply FRU by referencing the Illustration below.

    Figure 3. Z840 Component Location

  4. Pull out the old power supply from the Host computer, replace the z840 power supply with new one.

    Figure 4. Power Supply Removal

  5. Reinstall all removed components, and reconnect any cables that have been disconnected.

3 Restore the Console

Procedure

  1. Remove LOTO from console.
  2. Reinstall the host computer side cover.

4 Finalization

Procedure

  1. Confirm Host computer powers up when console power is turned on.
  2. Perform the Functional Checks → System Scanning Test instructions from the procedure list.