- Topic ID: id_17423188
- Version: 4.0
- Date: Aug 10, 2021 9:56:50 PM
Open Console (Z840) Power Supply Replacement
Prerequisites
Overview
This procedure shall be followed when replacing the Power Supply in the Z840 computer.
1 Preparation
Procedure
- Shutdown system. Select one of the following methods to Power
OFF the Console:
-
If Applications are up, click on the [Shut Down] button on desktop display and select [Shutdown].
-
If Applications are down, open a Terminal Window. Type: halt , then press ENTER.
-
When halt command has finished, power Off the console at the front panel switch.
-
- Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.
2 Power Supply Replacement
Procedure
- Remove the side access panel by loosen three screws and pull
out the latch to release the computer left side cover.
Figure 1. Side Access Panel Removal

- Remove the following components from the host computer (Z840):
-
Expansion Card Support
-
Airflow Guide
Figure 2. Z840 Airflow Guide and Expansion Card Support

-
- Locate Power Supply FRU by referencing the Illustration below.
Figure 3. Z840 Component Location

- Pull out the old power supply from the Host computer, replace
the z840 power supply with new one.
Figure 4. Power Supply Removal

- Reinstall all removed components, and reconnect any cables that have been disconnected.
3 Restore the Console
Procedure
- Remove LOTO from console.
- Reinstall the host computer side cover.
4 Finalization
Procedure
- Confirm Host computer powers up when console power is turned on.
- Perform the Functional Checks → System Scanning Test instructions from the procedure list.