- Topic ID: id_17423195
- Version: 4.0
- Date: Aug 10, 2021 9:56:49 PM
Open Console (Z840) FDIP Card Replacement
Prerequisites
Overview
This procedure shall be followed when replacing the FDIP Card in the host computer (Z840).
1 Host Computer Removal
Procedure
- Shutdown system. Select one of the following methods to Power
OFF the Console:
-
If Applications are up, click on the [Shut Down] button on desktop display and select [Shutdown].
-
If Applications are down, open a Terminal Window. Type: halt , then press ENTER.
-
When halt command has finished, power Off the console at the front panel switch.
-
- Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.
2 FDIP Card Replacement
Procedure
- Remove the side access panel by loosen three screws and pull
out the latch to release the computer left side cover.
Figure 1. Side Access Panel Removal

- Remove the Remove the following components from the host computer
(Z840):
-
Expansion Card Support.
-
Airflow Guide
Figure 2. Z840 Airflow Guide and Expansion Card Support

-
- Locate DIP Card FRU by referencing the Illustration below.
Figure 3. Z840 Component Location

- Remove the Optic fiber cable from computer rear side.
- Replace the existing FDIP card (Slot 1) with new one.
- Reinstall all removed components, and reconnect any cables that have been disconnected.
3 Restore the Console
Procedure
- Remove LOTO from console.
- Reinstall the host computer side cover.
4 Finalization
Procedure
- Confirm Host Computer powers up when console power is turned on.
- Perform the Functional Checks → System Scanning Test instructions from the procedure list.