• Topic ID: id_17423156
  • Version: 4.0
  • Date: Aug 10, 2021 9:56:56 PM

Host Computer (Z840) FDIP Card Replacement

Prerequisites

Overview

This procedure shall be followed when replacing the FDIP Card in the host computer (Z840).

1 Host Computer Removal

Procedure

  1. Shutdown system. Select one of the following methods to Power OFF the Console:
    • If Applications are up, click on the [Shut Down] button on desktop display and select [Shutdown].

    • If Applications are down, open a Terminal Window. Type: halt , then press ENTER.

    • When halt command has finished, power Off the console at the front panel switch.

  2. Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.
  3. Remove Front and Top covers. Refer to the following procedure.

    Refer to Replacement → Console → Console Cover Removal and Installation

  4. Remove the host computer from console chassis. Refer to the following procedure.

    Refer to Replacement → Console → Host Computer (Z840) Replacement

2 FDIP Card Replacement

Procedure

  1. Open the host computer side access panel.
  2. Remove the Expansion Card Support.

    Figure 1. Z840 Airflow Guide and Expansion Card Support

  3. Replace the existing FDIP card with new one.

    Figure 2. Z840 Component Location

  4. Install the Expansion Card Support and close the Side Access Panel.

3 Restore the Console

Procedure

  1. Install Host computer into Console chassis. Refer to the following procedure.

    Refer to Replacement -> Console -> Host Computer (Z840) Replacement

  2. Reconnect all cables removed earlier to the Z840 computer.
  3. Install Console covers. Refer to the following procedure.

    Refer to Replacement → Console → Console Cover Removal and Installation

  4. Remove LOTO on console.

4 Finalization

Procedure

  1. Perform the Functional Checks → System Scanning Test instructions from the procedure list.