- Topic ID: id_11038918
- Version: 4.0
- Date: Apr 22, 2019 12:56:10 AM
Detailed Calibration
Prerequisites
This document provides the necessary steps to detailed calibrate the system.
1 Setup
Procedure
- Move table to longitudinal home position.
- Attach phantom holder to gantry end of cradle.
- Mount phantom to phantom holder.
- Place bubble level on top of phantom, positioned along Z-axis. Adjust phantom holder back knob until phantom is level front to back.
2 Calibration
Procedure
- Click SCANNER UTILITIES.
- Click DETAILED CALIBRATION. Gantry Balance is performed.
- Verify all FOCAL SPOT are selected (highlighted).note:
All KV, SFOV, and SLICE COLLIMATION are forced on and not selectable.
- Click CONFIRM.
- Click OK in Attention window.
- Press START on SCIM when it flashes. Air Calibrations are performed.
- Center phantom.note:
Setup and centering must be completed and calibration started in less than 30 minutes, otherwise required to start over.
- Turn on laser alignment lights.
- Move table into gantry until axial internal alignment lights are aligned with phantom's axial reference mark.
- Adjust phantom holder side knob until sagittal internal alignment lights are aligned with phantom's sagittal reference mark.
- Elevate table until coronal internal alignment lights are aligned with phantom's coronal reference mark.
- Set Internal Landmark.
- Move table out of gantry 80mm.note:
Axial internal alignment lights should be in water area of phantom.
- Elevate table 40mm.
- Click CONFIRM.
- Press START on SCIM when it flashes. Phantom Centering is performed.
- Adjust phantom position if results are out of specification
range.note:
Phantom center specification is ± 1.0mm.
- Repeat steps [h] – [j] until phantom is centered.
- Click CONTINUE when phantom is centered.
- Press START on SCIM when it flashes. Water Phantom Calibrations (Small) are performed.
- Elevate table 40mm.
- Center phantom.
- Click CONFIRM.
- Press START on SCIM when it flashes. Phantom Centering is performed.
- Adjust phantom position if results are out of specification
range.note:
Phantom center specification is ± 1.0mm.
- Repeat steps [a] – [c] until phantom is centered.
- Press CONTINUE when phantom is centered.
- Press START on SCIM when it flashes. Water Phantom Calibrations (Large) are performed.
- Click OK in Attention window.
- Click QUIT in Scanner Utilities window.
3 Finalization
Procedure
- Fast Calibration
- Save System State