• Topic ID: id_17423388
  • Version: 6.0
  • Date: Jan 20, 2020 8:31:54 PM

Collimator Calibration

Prerequisites

Overview

This document provides necessary steps to detailed calibrate the system.

Procedure

  1. Move table to longitudinal home position.
  2. Click SCANNER UTILITIES.
  3. Click COLLIMATOR CALIBRATION.
  4. Click OK in Attention window.
  5. Press START on SCIM/GSCB when it flashes.
  6. Click OK in Attention window.
  7. Click QUIT in Scanner Utilities window.

Finalization

  1. Detailed Calibration
  2. Fast Calibration
  3. Save System State