Band Test
Prerequisites
Overview
This document provides the necessary steps to test for band artifacts.
Procedure
- Perform Center Phantom.
- Set Internal Landmark.
- Select NEW PATIENT.
- Type the following entries in the listed Patient Information fields:
- Patient ID: Service
- Patient Name: Band Test
- Select Service from Protocol Selection, Anatomical Selector.
- Select MANUFACTURING, IMGSER 20CMQA.
- Select Series 2, Series Description: QA3 Heli/0.51/120kV/135mA/1s, by clicking NEXT SERIES twice.
- Click CONFIRM.
- Press MOVE TO SCAN on SCIM when it flashes.
- Press START on SCIM when it flashes.
- Click END EXAM.
- If PPS Response window is displayed, click COMPLETE.
- Select ImageWorks desktop.
- Select scanned examination from List/Select browser.
- Click VIEWER and visually review all images for band artifacts.
- For images depicting bands, perform the following steps.
- Select SERVICE desktop and move Service desktop to left monitor.
- Select IMAGE QUALITY from Service desktop.
- Select IMAGE ANALYSIS and move Image Analysis window to left monitor.
- Select image from right monitor List/Select browser.
- Click MANUAL radio button on Image Analysis window.
- Click MANUAL selection button, BAND.
- Click ACCEPT.
- Follow instructions in Image Analysis window lower right pane, to size Band ROI and analyze.
- From Report window, if Avg. Ref. Band Factor is out of specification range +/-2.8, perform appropriate corrective action.
- Click [DISMISS] in Report window.
- Click QUIT in Image Analysis window.
Finalization
No finalization steps.