| Rev |
Date |
Reason For Change |
| 15 |
May 11, 2022 |
Replacement > Table (GT 1700)
Add "Table Speaker Replacement"
> Lite Table
Add "Table Speaker Replacement"
|
| 14 |
July 23, 2021 |
Installation > Option
Add "Image Protocol Manager 1.0 Installation"
|
| Software > Software Installation Procedure (LFC)
Add "(20BW25.5) Service Pack Installation Procedure"
|
| 13 |
April 06, 2021 |
Align, Setup, Calibrations >Console
Update "System Configuration- 15BW21.x and 21BW02.x"
|
| Software
>Software Installation Procedure (LFC)
Update "(15BW21.x and 21BW02.x) Load From Cold"
|
| Replacement > Gantry > High Voltage
Update "HEMIT Tank Replacement" per SPR HCSDM00651779
|
| 12 |
Sep 21, 2020 |
Installation Update Pre-Installation manual 5472000-1EN to Rev 7
>Option
Add "Table Protection Pad Installation Manual"
|
| Align, Setup, Calibrations >Console
Add "Data Privacy (EA3 User Authorization) Configuration"
Update "System Configuration- 16BW45.x or later"
Add "Product Locator"
|
| Software Add "System Time Adjustment"
>Software Installation Procedure (LFC)
Update "(18BW26.x and 20BW25.x) Load From Cold"
|
| Theory Add "Liquid Bearing Tube Service Tool Impact"
Add "Liquid Bearing Tube Rotor Control Thermal Guidance for Hot landing prevention"
>Gantry
Update "Performix Plus tube theory with liquid bearing rotor management"
|
| Troubleshooting >Console (Open Console)
Add "Recon GPU Card Troubleshooting"
>Console (NIO Console)
Add "Recon GPU Card Troubleshooting"
>X-Ray Generation
Add "Rotor Control Tool with LB tube - GUI (20BW25.x or later)"
Add "Rotor Control Tool with LB tube - Command"
|
| Part List Update "Gantry Parts List"
Update "Table Parts List"
|
| Replacement >Console (OpenOC)
Update "Open Console (Z840) Recon GPU Card Replacement"
>Console (NIO OC) > NIO64 Console_Z840 Host Computer
Update "Host Computer (Z840) Recon GPU Card Replacement"
>Table
Add "Cradle Extender Latch Replacement"
>GT1700V Table
Add "GTCB Cover Replacement"
Add "GT Top Side Cover Bracket"
|
| 11 |
Feb 06, 2020 |
Updated Installation manual 5472001-1EN to Rev 9
|
| Safety Add "Liquid Bearing Tube Safety"
|
|
Software
>Software Installation Procedure (LFC)
Update "(18BW26.x) Load From Cold"
Update "(16BW45.x / 17BW34.x) Load From Cold"
Update "(15BW21.x) Load From Cold - Optima CT680 Series"
Update "(13HW38.x) Load From Cold - Optima CT680 Series"
|
| 10 |
Nov 06, 2019 |
Software
>Software Installation Procedure (LFC)
Updated “(18BW26.x) Load From Cold, (16BW45.x /17BW 34.x) Load From Cold, (15BW21.x) Load From Cold, (13HW38.x) Load From Cold ” per SPRHCSDM00573443
|
| Replacement
> Gantry > Tube
Updated “X-Ray Tube Replacement"
|
| 9 |
Sep 06, 2018 |
Updated Installation manual 5472001-1EN to Rev 8
|
|
Software
>Software Installation Procedure (LFC)
Updated “(18BW26.x) Load From Cold”
|
| 8 |
July 17, 2018 |
Installation
Updated Pre-Installation manual 5472000-1EN to Rev 6
Updated Installation manual 5472001-1EN to Rev 7 |
|
Software
>Software Installation Procedure (LFC)
Added “(18BW26.x) Load From Cold”
|
|
Theory
>Console
Added “Open Console with Z840 Theory”
|
|
System Diagrams
Added “Optima CT680 System Block Diagram with Z840”
Added “Open Console Interconnects” and “Open Console AC Box Outlet Diagram” |
|
Troubleshooting
Added Open Console Folder for Open Console related contents
|
|
Part List
Updated Gantry Part List
Added Open Console Part List
|
|
Replacement
Added Open Console Folder for Open Console related parts replacement procedure
|
|
Planned Maintenance and Tasks
Change the PM contents according to global service team requirement
|
| 7 |
April 28, 2018 |
Replacement
>Gantry
Added “Gantry Stationary Power Supply 120VAC Cable Replacement” |
| 6 |
Sep 04, 2017 |
Installation
Updated Pre-Installation manual 5472000-1EN to Rev 5
Updated Installation manual 5472001-1EN to Rev 6 |
|
Software
>Software Installation Procedure (LFC)
Updated “(16BW45.x / 17BW34.x) Load From Cold”
|
| 5 |
May 05, 2017 |
Readme
Updated “Secure Service Access Generation 2 (SSA2) CSD Tool List Reclassification”
|
|
Installation
Updated Pre-Installation manual 5472000-1EN to Rev 4
Updated Installation manual 5472001-1EN to Rev 5 |
|
Align, Setup Calibration
>Console
Added “ Data Privacy (EA3 User Authorization) Configuration (16BW45.x or later)” Added “How to add Body-Part Examined Tag (0018,0015) to Image Header”
Added “NIO64 (Z840) CMOS Battery Replacement, BIOS Settings and ROM Flashing”
Added “LCD Video Monitor Setup for Z840” Added “System Configuration - Optima CT680 (16BW45.x or Later)”
>Gantry Added “Configuring Gantry Tilt for Extender Removal” procedure |
|
Software
Added “Configure HIPAA procedure”
>Software Installation Procedure (LFC)
Added “(16BW45.x) Load From Cold”
Added “(16BW45.x) LFC Quick Guide” |
|
Theory
>Console
Added “(Z840) Console Theory” |
|
System Diagrams
Updated Optima CT680 Functional Interconnects
|
|
Troubleshooting
>Console
Added “NIO64 (Z840) Computer Troubleshooting” Added “RAID Performance Measurement”
Added “Recon Performance Verification Procedure” Added “Aegis SSA Key Workaround for older CT systems Software”
Added “Troubleshooting Information for Image Browser Blue line issue” |
|
Part List
Added Z840 part into NIO64 Console Parts List
|
|
Replacement
Added new folder for z840 FRU part replacement
> Gantry >DAS (Merc40) Added “Rotating 48VDC 1kW Power Supply Replacement from Original 1.8kW Power Supply” procedure |
|
Planned Maintenance and Tasks
Added “Planned Maintenance Schedule W” according to global service strategy
|
| 4 |
May 20, 2016 |
Installation
Updated Installation manual 5472001-1EN to Rev 4
Added Installation Forms |
|
Align, Setup Calibration
>System
Added “RTP Alignment Test Procedure” |
|
Part List
Updated Gantry Parts List
Updated NIO64 Console Parts List |
|
Replacement
Updated “Certified Component Replacement Retest Matrix” and “Certified Component Verification Form e4879”
|
|
Planned Maintenance and Tasks
Deleted 4 Times PM / Year according to global service strategy
|
| 3 |
Sep 07, 2015 |
Installation
Updated Pre-Installation manual 5472000-1EN to Rev 3
Updated Installation manual 5472001-1EN to Rev 3
>Option Added “Cardiac Monitor (IVY) Installation”
|
|
Align, Setup Calibration
>Console
Added “Data Privacy (EA3 User Authorization) Configuration (15BW21.x or later)”Added “Dose Check Management and EA3 Configuration (15BW21.x or later)”
Added “GSCB Configuration and Intercom Setup”
> Gantry->Enclosure: Added “Backlight brightness adjusting tool for Gantry Control Panel”
> System:
Added “ System Alignment Procedure”> Table->GT1700V:
Added “ Table Load Height Setting” |
|
Software
Added “SSA License File Installation”
Added “Unmount SSA key procedure”
> Software Installation Procedure (LFC): Added “15BW21.x Load From Cold - Optima CT680”
Added “(13HW38.x / 15BW21.x) LFC Quick Guide” Updated “(13HW38.14) Service Pack Installation Procedure”
|
|
Theory
>Gantry->DAS and Detector
Updated “CDAS Theory” according to DOC1528609_r2 |
|
System Diagrams
Updated “Optima CT680 System Block Diagrams”
Updated “Gantry Cable Interconnect” Added “Merc40 Detector Heater Control” diagram
Added “Merc40 Detector Plenum Fan Control” diagram |
|
Troubleshooting
>Gantry
Updated “Slipring Brush Disconnects Troubleshooting”>Image Quality
Added “Z alignment Cal. Troubleshooting”>Table
Added “Precise T/G alignment” |
|
Part List
Updated Gantry Parts List
Updated Table Parts List Updated PDU Parts List
Updated Misc Item Parts List |
|
Replacement
Added “NGPDU-81 ABB AF116 Kxg Contactor Replacement”
>Console
Added “HP PC Battery Replacement”
>Gantry->DAS (Merc40) Added “Merc40 DAS-Detector Assembly Replacement Procedure”>Table->GT1700V
Added “RIM IMS Cover Set Replacement”PDU Added “PDU Covers Remove/Install”
Added “PDU Spare Part Replacement Procedure”
|
|
Planned Maintenance and Tasks
Added new folder for 3 Times PM / Year
|
| 2 |
April 09, 2014 |
Installation
Updated Pre-Installation manual 5472000-1EN to Rev 2
Updated Installation manual 5472001-1EN to Rev 2
>Option Added “Long Cable Kit Installation (Dir 5456816–1EN)”
|
|
Align, Setup, Calibrations
> Console: Updated “Z820 Host PC BIOS Setup”
|
|
Functional Checks
Added “Enclosure Leakage (Patient Touch) and System Chassis Leakage Tests”
> Gantry: Added “zalignmentWizard User Instructions (Merc40)” |
|
Software
> Software Installation Procedure (LFC): Updated “13HW38.x Load From Cold - Optima CT680”
Added “(13HW38.14) Service Pack Installation Procedure” Updated “ Install Software Options”
|
|
Theory
> Gantry: Added “Gantry Thermal Theory (Merc40)”
> Gantry->DAS and Detector: Added “CDAS Theory” Added “Merc40 Detector Theory”
|
|
System Diagrams
Updated “Optima CT680 System Block Diagrams”
|
|
Troubleshooting
> Console: Updated “NIO64 Console Diagnostics”
> Diagnostics: Added “DAS Tools (Merc40)” Added “Scan Data Path Diagnostics (Merc40)”
> Gantry: Added “Gantry Thermal Control Troubleshooting (Merc40)”
> Image Quality: Added “Scan Analysis (Merc40)”
|
|
Part List
Updated Gantry Parts List
|
|
Replacement
Added “Gantry Heater Fuse Holder Replacement”
> Gantry->DAS: Added related Merc40 component replacement procedure
> Gantry->Enclosure: Added “Gantry Top Fan replacement” procedure
Added “Gantry Heater replacement” procedure
> Gantry->ORP and TGPG: Added “ORP Replacement (Merc40)” procedure Added “TGPG Replacement” procedure
Added “TGPG Power Supply Replacement” procedure |
|
Planned Maintenance and Tasks
Updated “Planned Maintenance Schedule A”
Updated “Planned Maintenance Schedule B” Updated “Planned Maintenance Schedule C”
Updated “Planned Maintenance Schedule D” |
| 1 |
Dec 20, 2013 |
Initial Release for KL64 Saturn MV |