Gantry Tilt Zero Check
-
Navigate to the Common Service Desktop.
-
Click the tab.
-
Click the (near the bottom of the menu list). This launches the tool.
-
Select the screen.
-
Select the tab.
-
View the Tilt Angle section.
-
At the gantry:
-
Tilt the gantry to +10 degrees.
-
Press both tilt buttons until the tilt stops at zero.
-
-
At the console, view the Tilt Angle data section. The reading should be 0 ± 0.14.
-
If the reading is not correct:
-
Repeat Steps 1 - 8 one more time.
-
If the reading still is not correct, adjust the Tilt Pot. (For procedures, see Gantry Tilt Pot Adjustment.)
-
-
Record the results on the PM Report Form.