Liquid Bearing Tube Rotor stop procedure
Prerequisites
1 Tube Rotor Stop Procedure - Advanced
Procedure
- Start Rotor Control Tool under Diagnostic tab from CSD menu.
- Check the current rotor status. If it’s not Stop, then check Stop and press Accept button to stop the rotor.
Figure 1. Window - Rotor Control Tool

- Confirm that the Tool Status shows Rotor is coasting or stopped and press Dismiss button to close the tool.
Figure 2. Window - Tool Status

2 Finalization
No finalization steps.