• Topic ID: id_11038811
  • Version: 4.0
  • Date: Nov 27, 2020 2:13:28 AM

Slipring Receiver (850 Mbaud) Replacement

Prerequisites

This procedure describes the Slipring Receiver replacement.

Procedure

  1. Move table to its lowest elevation.
  2. Remove rear gantry cover Gantry Rear Cover Removal and Re-install.
  3. Remove J1 and J2 cables.
  4. Remove hex screws holding the receiver to the antenna.

    Figure 1. 850 MBaud Slipring Receiver

  5. Replace receiver.
  6. Verify antenna gap and alignment adjustments.

Finalization

  1. Hardware Reset
  2. Acquire 10 scouts: (120kV/40mA, 1000mm table movement)
  3. Acquire 100 axials: (120kV/80mA, 0.5 sec. Scan)
  4. Acquire 1 helical: (120kV/40mA, 30 sec. Scan),
  5. Acquire 10 axial scans: (120kV/400mA, 4 sec. Scan)
  6. Verify NO increase in corrected or uncorrected FEC errors.
  7. Reset DIP Stats: Perform the following procedure to reset cumulative DIP Stats (shown on the Common Service Desktop Home Page):
    1. Open a Unix shell.
    2. Type: cd /usr/g/service/log
    3. Type: mv dip.stats dip.stats.old
    4. Shutdown and restart the system. This will reinitialize the dip.stats file.
    5. Save System State.