• Topic ID: id_11038923
  • Version: 5.0
  • Date: Jan 20, 2020 8:36:42 PM

Save Restore Generator Runtime Parameters

Prerequisites

Overview

This document defines the procedures to save and restore the Generator Runtime Parameter values, which are used during Generator failure identification and troubleshooting.

1 Save Generator Runtime Parameters

Procedure

  1. Select DIAGNOSTICS from Service Desktop.
  2. Select GENERATOR TOOL – JEDI, or X-RAY GENERATION, GENERATOR TOOL – JEDI.
  3. Click SAVE RUNTIME PARAMETERS.
  4. Click OK in Attention window.
  5. Click OK in Save Runtime Parameters window.
  6. Click DISMISS in Generator Tool - JEDI window.
  7. Save System State.

2 Restore Generator Runtime Parameters

Procedure

  1. Select DIAGNOSTICS from Service Desktop.
  2. Select GENERATOR TOOL – JEDI, or X-RAY GENERATION, GENERATOR TOOL – JEDI.
  3. Select RESTORE RUNTIME PARAMETERS.
  4. Click OK in Attention window.
  5. Click OK in Restore Runtime Parameters window.
  6. Click DISMISS in Generator Tool - JEDI window.

3 Finalization

No finalization steps.