Recon Performance Verification Procedure

Prerequisites

Overview

This procedure describes how to verify the recon performance speciation which is provided in the Product Datasheet.

1 Acquire Images

Procedure

  1. Place QA phantom on cradle.
  2. Set Internal Landmark.
  3. Select NEW PATIENT.
  4. Type the following entries in the listed Patient Information fields:
    1. Patient ID: Service
    2. Patient Name: Recon Performance Test
  5. Select Service from Protocol Selection, Anatomical Selector.
  6. Select Recon Performance Test.

    (For System within Enhance Recon Option) the scan protocol is shown as below:

    (For System without Enhance Recon Option) the scan protocol is shown as below:

  7. Click CONFIRM.
  8. Press MOVE TO SCAN on Scan Control Box when it flashes.
  9. Press START on Scan Control Box when it flashes.
  10. By using a stop watch, measure the time between the first image display and the last image display.

2 Calculation of Reconstruction Time

Procedure

  1. Calculate the reconstruction time (fps) per the following formula:

    Reconstruction Time (fps) = (total number of images -1) / (time between the first image display and the last image display)

    Specification: ≥30 fps

    note: The specification may change by new software release. Confirm with the latest product data sheet.

3 Finalization

No finalization steps.