Prerequisites
This procedure shall be followed when replacing the power supply
in the Z820 computer.
Procedure
- Select one of the following methods to Power OFF the Operator
Console:
- If Applications are up, click on the Shut Down button and select Shutdown.
The Operator Console monitor will display a 'Power Down' message
when it is acceptable to power OFF the Operator Console.
- If Applications are down, open a Unix Shell. Type: halt.
- Power OFF the console at the front panel switch.
- Remove the left side cover from the console.
- Remove the side access panel.
- Pull out the old power supply from the Host computer, replace
the z820 power supply with new one.
Figure 1. Power Supply Removal
- Reinstall the side access panel.
Finalization
- Refer to System Scanning Test to confirm proper operation.
- Reinstall the console covers.