| 260132300 |
DAS CAN Communication Object Initialization Error, Object Id:
%d; Error Code: %d |
| 260132301 |
Unable to open remote CAN node for communication. Attempted
To Open CAN Node Id: %d; Received Error Code: %d |
| 260132302 |
Unable to open communication object, Object Id: %d; Error Code:
%d |
| 260132303 |
Unable to close communication object, Object Id: %d; Error
Code: %d |
| 260132304 |
Scan ABORT -- DAS Has Detected An Error. Communication object
(OID) %d; detected receive error %d; MessageCode: xxx. |
| 260132309 |
Scan ABORT -- DAS Has Detected An Error. Unable to send message
%d to CAN node: %d object: %d. Communication error %d detected. |
| 260132310 |
Not all data acquisition system tasks have started. Task initialization
flags are: xxx |
| 260132311 |
Unable To Create Data Acqusitition System Task. Operating System
Error Code: %d |
| 260132316 |
Scan ABORT -- DAS Has Detected An Error. Operating system ERROR
%d detected. |
| 260132317 |
Scan ABORT -- DAS Has Detected An Error. Memory Allocation
failure. Unable to allocate %d bytes of memory. |
| 260132320 |
Scan ABORT -- DAS Has Detected An Error. DAS was receiving
Acquisition Triggers when a First View Azimuth message was received.
(Scan State: %b) Trigger Frequency: %d Hz. (Triggers: %d Views - Odd:
%d, Even: %d). |
| 260132321 |
Scan ABORT -- DAS Has Detected An Error. DAS was expecting
Acquisition Triggers when a First View Azimuth message was received.
(Scan State: %b) |
| 260132322 |
Scan ABORT -- DAS Has Detected An Error. Unexpected Acquisition
Trigger received. (State: %b Command Level: %b ScanState: %b) |
| 260132324 |
Scan ABORT -- DAS Has Detected An Error. Acquisition hardware
(DCB) has detected an Unacceptable Amount of Acquisition Trigger Jitter.
(Trigger Frequency: %d hertz). (State: %b Command Level: %b ScanState:
%b) |
| 260132329 |
Scan ABORT -- DAS Has Detected An Error. Data acquisition control
board, command table error. Command %d has a NULL function pointer.
Command level is: %d. |
| 260132330 |
Unable To Notify DCB State Machine Of Scan Event. Scan Event:%d;
Error code:%d. |
| 260132331 |
DAS Has Detected An Error In Download Operation. Invalid download
code received. Received code: %d |
| 260132332 |
DAS Has Detected An Error In Download Operation. Download request
is too large for DAS buffers. Request was for %d bytes. Maximum buffer
size is %d bytes. |
| 260132333 |
DAS Has Detected An Error In Download Operation. Download operation
request codes do not match. Base download request code is: %d. Download
request code embedded in operation is: %d. |
| 260132334 |
DAS Has Detected An Error In Download Operation. Downloaded
image corrupted. Base download request code is: %d. |
| 260132345 |
DAS Has Detected An Error In Download Operation. FLASH memory
write error. |
| 260132346 |
Invalid Trigger Frequency Rx Parameter Received. Requested:
%d hertz. Valid range is: %d to %d hertz. |
| 260132347 |
Invalid DAS Gain Rx Parameter Received. Data Channel Gain:
%d. Z Lower Channel Gain: %d. Z Upper Channel Gain: %d. |
| 260132350 |
Rx Selected An Invalid View Format Type. Data Acquisition Hardware/Firmware
Does Not Support the Requested View Format. Check DAS configuration.
View Format Requested: %x |
| 260132352 |
Scan ABORT -- DAS Has Detected An Error. Error detected accessing
data acqusition system hardware. |
| 260132354 |
Scan ABORT -- DAS Has Detected An Error. Scan Setup Errors
Prevent The Data Acquisition System From Being Able To Scan. (See
Previous Log Entries.) |
| 260132355 |
DAS Selftest Command Table Error Prevents Execution Of Selftest.
Selftest command %d has a NULL function pointer. |
| 260132364 |
Scan ABORT -- DAS Has Detected An Error. Errors detected while
Reconfiguring the Acquisition system for Scan Data Views. |
| 260132368 |
Null Statemachine Control Structure Pointer Received. |
| 260132384 |
Scan ABORT -- DAS Has Detected An Error. The DAS has received
an invalid command. Command %d is not recognized by the DAS. (DAS
CommandState: %b, CommandLevel: %b, ScanState: %b) |
| 260132385 |
Scan ABORT -- DAS Has Detected An Error. The DAS has received
an out-of-sequence command. Command %d cannot be accepted when the
DAS is in the State: %b, CommandLevel: %b, ScanState: %b. |
| 260132390 |
DCB failure detected while writing to %b. (Address: xxx) Expected:
xxx, Actual: xxx |
| 260132405 |
DCB detected Coverter Board Fault Line Asserted. No Detected
Configured Converter Board(s) Asserted it. |
| 260132410 |
DAS Rx Message Is Corrupt. Expected and Actual Rx Message CRC's
Do Not Match. |
| 260132416 |
Self test TIMEOUT occurred on Internal Trigger Scan. |
| 260132419 |
Detector Temperature Has Returned To Normal. |
| 260132421 |
DCB Current State Level is INVALID. Current State Level: %d. |
| 260132428 |
The DAS did NOT receive any Triggers. The scan will be Aborted. |
| 260132444 |
Scan ABORT -- DAS Has Detected An Error. DAS was expecting
EXPCMD Interrupt when a Start Internal Triggers Acquistion message
was received. (Scan State: %b) |
| 260132445 |
Scan ABORT -- DAS Has Detected An Error. DAS was NOT expecting
Stop Internal Triggers message during a Internal Triggers Acquistion.
(Scan State: %b) |
| 260132455 |
Beam Tracking Calculated A Focal Spot Size. That Has Exceeded
The WARNING Tolerance. Tolerance: %6d umf, Caclulated Size: %6d umf,
Nominal Size: %6d umf |
| 260132457 |
Beam Tracking Control Loop Error Exceeded The WARNING Tolerance.
Tolerance: %6d umd, Control Error: %6d umd, Beam Side: %b |
| 260132458 |
Beam Tracking Control Loop Error Exceeded The ABORT Tolerance.
Tolerance: %6d umd, Control Error: %6d umd, Beam Side: %b |
| 260132461 |
Beam Tracking Detected Signal Corruption. mA Level: %ld, Z
Channels: %ld, Offset Vector: %ld |
| 260132470 |
Unable to Notify Detector Manager Command to the Detector Manager
Task. Error code %d received. |
| 260132486 |
Scan ABORT -- DAS Has Detected An Error. Acquisition hardware
(DCB) has detected an INVALID Number of Views versus Triggers. (Trigger
Frequency: %d Hz). (Triggers: %d, Views - Odd: %d, Even: %d). (State:
%b, Command Level: %b, ScanState: %b) |
| 260132487 |
Scan ABORT -- DAS Has Detected An Error. The DCB has detected
IFB First View Timeout error(s). IFB First View: Timeout: %u nsec.
(IFB First View Timeout Interrupt) |
| 260132488 |
Scan ABORT -- DAS Has Detected An Error. The DCB has detected
IFB Subsequent View Timeout error(s). IFB View: Timeout: %u nsec.
View Number: %u. (IFB Subsequent View Timeout Interrupt) |
| 260132489 |
Scan ABORT -- DAS Has Detected An Error during Acquisition.
(Trigger Frequency: %d hertz). (Triggers: %d). (State: %b Command
Level: %b ScanState: %b) |
| 260132490 |
An illegal summation cell was specified in the DAS Rx provided
by the host. Row: %d Cell: %u. |
| 260132491 |
Thread object of class \"%b\" unexpectedly destructed. |
| 260132494 |
Unable to create the DCB Message Queue object creation status:
%d, queue size: %d entries |
| 260132495 |
Unable to create the DCB Message buffer Pool object creation
status: %d, creating %d buffers of %d bytes each |
| 260132496 |
DAS Has Detected An Error In Download Operation. Expected the
download message state to be: %b. Found the message state instead
to be: %b. |
| 260132502 |
Scan ABORT -- DAS Has Detected An Error. The DCB has detected
IFB DPA Lost Lock error(s). View Number: %d. |
| 260132503 |
Scan ABORT -- DAS Has Detected An Error. The DCB has detected
IFB Checksum error(s). View Number: %d. |
| 260132504 |
Scan ABORT -- DAS Has Detected An Error. The DCB Has Detected
An Interface Board Has failed To Generator The Correct Number of Views.
Interface Board:%d, DCB Trigger Count: %d, IFB Trigger Count: %d,
IFB Data Count: %d, IFB View Ouput: %d |
| 260132509 |
Scan ABORT -- DAS Has Detected An Error. The DCB has detected
Ifb Rx FIFO Fill error(s). Skew 0: xxx, Skew 1: xxx, Skew 2: xxx,
Skew 3: xxx |
| 260132510 |
Scan ABORT -- DAS Has Detected An Error. The DCB detected IFB
view data checksum error. |
| 260132516 |
Communication To The Command Fan Control Board Has Not Been
Established. Scanning Is Disabled Until The Thermal Status of the
DAS is Known. |
| 260132517 |
Communication To The Detector Heater Control Board Has Not
Been Established. Scanning Is Disabled Until The Thermal Status of
the DAS is Known. |
| 260132521 |
Hold Wait For Setup Done State. |
| 260132522 |
Pending Rx Wait For Setup Done State. |
| 260132523 |
Attempt to dereference an invalid iterator. |
| 260132524 |
The DAS Has Detected Missing Triggers During This Acquisition.
This Typically Occurs Due to Slip-Ring Brush Disconnect. Call Service
For Maintenance. Trigger Count: %d |
| 260132525 |
Platform Trigger Count |
| 260132526 |
Scan ABORT -- DAS Has Detected An Error. Specified scan will
overflow the IFB's internal buffers. # Triggers (views): %u, Trigger
frequency: %u Hz, IFB Buffer size: %u MBytes |
| 260132527 |
DAS has Detected Issue with mA/kV Packets. Triggers Received:
%u, Scorp Packets received: %u, Scorp Checksum errors: %u |
| 260132528 |
DAS has Detected Optical high speed link dropped packets. OpticalKError:
%u, OpticalRdError: %u, OpticalRsdCwDecFail: %u, OpticalRsdCwError:
%u, OpticalLocked: %u |
| 260132529 |
Scan ABORT -- DAS Has Detected An Error. 0 - Check Passed,
1 - Check Failed, 255 - The check was not done, IFBs Ready: %u,
Acq Setup Failed: %u, Odd No of PreTrigg: %u |
| 260132530 |
Scan ABORT -- DAS Has Detected An Error. Did not get the minimum
number if offsets Min no of offset triggers required: %u, No of offset
triggers received: %u |
| 260132531 |
DAS has not detected any IFB. 0 - DAS will Abort all scans,
1 - DAS will Scan in simulated Mode, DAS Status: %u |
| 260132532 |
Setup Scan Sequence Error. Error in step: %b. |
| 260132556 |
Incorrect Power Module ID was requested. Requested ID: %d,
Available ID: 1 or 2 |
| 260132557 |
Over Temperature Condition of Power Module was Detected. PM1-1:
%d deg, PM1-2: %d deg, PM1-3: %d deg, PM1-4: %d deg, PM1-5: %d deg,
PM2-1: %d deg, PM2-2: %d deg, PM2-3: %d deg, PM2-4: %d deg, PM2-5:
%d deg |
| 260132558 |
At Least One of Temperature Sensors of Power Module is Broken
or Disconnected. The Status is below. 1 means alive. PM1-1: %d, PM1-2:
%d, PM1-3: %d, PM1-4: %d, PM1-5: %d, PM2-1: %d, PM2-2: %d, PM2-3:
%d, PM2-4: %d, PM2-5: %d |
| 260132559 |
Incorrect Temperature Sensor ID of Power Module was requested.
Requested Temp Sensor ID: %d, Requested Power Module ID: %d, Available
Temp Sensor ID: 1 - 5 |