• Topic ID: task_i3v_hyc_zhb
  • Version: 4.0
  • Date: Nov 8, 2021 5:12:29 AM

Z8G4 NIC Card Replacement

Prerequisites

Overview

This procedure shall be followed when replacing the NIC Card in a Z8G4 computer.

note: Existing software option key will become invalid by NIC Card replacement. Regeneration of the software option key will be required by eLicense.

1 Preparation

Procedure

  1. Shutdown system. Select one of the following methods to Power OFF the Console:
    • If Applications are up, click on the [Shut Down] button on desktop display and select Shutdown.

    • If Applications are down, open a Terminal Window. Type: halt , then press ENTER.

    • When halt command has finished, power Off the console at the front panel switch.

  2. Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.

2 NIC Card Replacement

Procedure

  1. Open and remove the module small side cover

    Figure 1. Remove module small side cover

  2. Locate NIC Card by referencing the Illustration below.

    Figure 2. Z8G4 Component Location

  3. Remove the LAN cable from computer rear side.
  4. Reinstall all removed components, and reconnect any cables that have been disconnected.

3 Restore the Console

Procedure

  1. Remove LOTO from console.
  2. Reinstall the module small side cover.

4 Finalization

Procedure

  1. Confirm Host computer powers up when console power is turned on.
  2. Perform the Functional Checks → System Scanning Test instructions from the procedure list.
  3. Regenerate software option key by eLicense and install to the system.