- Topic ID: task_i3v_hyc_zhb
- Version: 4.0
- Date: Nov 8, 2021 5:12:29 AM
Z8G4 NIC Card Replacement
Prerequisites
Overview
This procedure shall be followed when replacing the NIC Card in a Z8G4 computer.
note: Existing software option key will become invalid by NIC Card replacement. Regeneration of the software option key will be required by eLicense.
1 Preparation
Procedure
- Shutdown system. Select one of the following methods to Power OFF the Console:
-
If Applications are up, click on the [Shut Down] button on desktop display and select Shutdown.
-
If Applications are down, open a Terminal Window. Type: halt , then press ENTER.
-
When halt command has finished, power Off the console at the front panel switch.
-
- Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.
2 NIC Card Replacement
Procedure
- Open and remove the module small side cover
Figure 1. Remove module small side cover

- Locate NIC Card by referencing the Illustration below.
Figure 2. Z8G4 Component Location

- Remove the LAN cable from computer rear side.
- Reinstall all removed components, and reconnect any cables that have been disconnected.
3 Restore the Console
Procedure
- Remove LOTO from console.
- Reinstall the module small side cover.
4 Finalization
Procedure
- Confirm Host computer powers up when console power is turned on.
- Perform the Functional Checks → System Scanning Test instructions from the procedure list.
- Regenerate software option key by eLicense and install to the system.