Prerequisites
Overview
This procedure describes the Slipring Power Supply Assembly
Replacement.
Procedure
- Remove gantry covers as required.
| warning
- Potential for Shock
- Voltage may be present.
- Use proper lockout/tagout procedures before replacing components.
See safety menu of Service Document gui for appropriate procedures.
|
| -
-
-
Always turn OFF the HVDC before the 120 VAC. Turning OFF 120
VAC power before HVDC power can result in equipment damage.
|
- Turn OFF all three (3) gantry service
switches (Axial Drive, HVDC, 120VAC). Follow Lockout/Tagout procedures.
- Using the hex socket, remove two (2) cap screws from the back
side of supply located on the rear right fan assembly.
- Remove safety plate.
- Disconnect wiring.
- Replace the slipring power supply assembly.
Finalization
- Perform System Scanning Test (located in folder).