Recon Performance Verification Procedure

Prerequisites

Overview

This procedure describes how to verify the recon performance speciation which is provided in the Product Datasheet.

1 Acquire Images

Procedure

  1. Place QA phantom on cradle.
  2. Set Internal Landmark.
  3. Select NEW PATIENT.
  4. Type the following entries in the listed Patient Information fields:
    1. Patient ID: Service
    2. Patient Name: Recon Performance Test
  5. Select Service from Protocol Selection, Anatomical Selector.
  6. Select Recon Performance Test.
    note: For Discovery RT, there may not be a protocol of Recon Performance Test, just manually enter the scan parameters.
  7. Click CONFIRM.
  8. Press MOVE TO SCAN on Scan Control Box when it flashes.
  9. Press START on Scan Control Box when it flashes.
  10. By using a stop watch, measure the time between the first image display and the last image display.

2 Calculation of Reconstruction Time

Procedure

  1. Calculate the reconstruction time (fps) per the following formula:

    Reconstruction Time (fps) = (total number of images -1) / (time between the first image display and the last image display)

    note: The specification may change by new software release. Confirm with the latest product data sheet.

3 Finalization

No finalization steps.