- Topic ID: id_23554117
- Version: 2.0
- Date: Sep 26, 2020 10:13:47 PM
PDAS Power Supply Replacement
Prerequisites
Overview
1 Power Supply Removal
Procedure
- notice
- Position the table to its lowest position.
- Remove the gantry right side cover.
Refer to .
- notice
- Turn OFF Axial Enable, HVDC, and 120VAC switches on the Service Switch panel.
- Remove the gantry top and front covers.
- Position the DAS at the 12 o'clock position.
- Lock the gantry in position using the rotational lock.
- Disconnect all cables connectors on the fuse board and power supplier.
- Remove 6 x M8 hex screws on power supplier.
Figure 1. Remove the screws from Power supplier

- Pull the power supplier by the handle and put on an ESD pad.
![]() |
![]() |
2 Power Supplier Installation
Procedure
- Get new power supplier and slide into place.
- Secure power supplier by using 6 x M8 screws. (Torque: 19.2 N.m)
- Reconnect all cables connectors on the fuse board and power supplier.
- Disengage the rotational lock.
- Turn on Axial Enable, HVDC, and 120VAC switches.