OpenOC16 (Z840) DIP Card Replacement

Prerequisites

Overview

This procedure shall be followed when replacing the DIP Card in a OpenOC (Z840) computer.

1 Preparation

Procedure

  1. Shutdown system. Select one of the following methods to Power OFF the Console:
    • If Applications are up, click on the [Shut Down] button on desktop display and select [Shutdown].

    • If Applications are down, open a Terminal Window. Type: halt , then press ENTER.

    • When halt command has finished, power Off the console at the front panel switch.

  2. Apply LOTO.

2 DIP Card Replacement

Procedure

  1. Remove the side access panel by loosen three screws and pull out the latch to release the computer left side cover.

    Figure 1. Side Access Panel Removal

  2. Remove the following components from the host computer (Z840):
    • Expansion Card Support

    • Airflow Guide

    Figure 2. Z840 Airflow Guide and Expansion Card Support

  3. Locate DIP Card FRU by referencing the Illustration below.

    Figure 3. Z840 Component Location

  4. Remove the Optic fiber cable from computer rear side.
  5. Replace the existing DIP Card (Slot 1) with new one.
  6. Reinstall all removed components, and reconnect any cables that have been disconnected.

3 Restore the Console

Procedure

  1. Remove LOTO from console.
  2. Reinstall the console covers.

4 Finalization

Procedure

  1. Confirm Host Computer powers up when console power is turned on.
  2. Perform the Functional Checks → System Scanning Test instructions from the procedure list.