- Topic ID: id_23554263
- Version: 1.0
- Date: Oct 9, 2018 1:40:47 PM
Heat Exchanger Filter Replacement
Prerequisites
Overview
This procedure shows how to replace the filter.
Procedure
warning- Remove power to the gantry using proper Lockout / Tagout procedures.
- Remove side, top, and front gantry covers.
- Turn off all three service switches on the Service Switch Panel (AXIAL DRIVE ENABLE, HVDC ENABLE, 120VAC).
- Rotate gantry so that Heat Exchanger is at the 2 O'Clock position (tube at or just below 3 o'clock).
- Remove the screws holding the filter in place. (See Figure 1.)
Figure 1. Screw Locations

- Replace Filter. Be sure the lower edge of the filter engages the tabs
on the heat exchanger frame.note:
The filter can be cleaned with a vacuum, if it is not excessively soiled.
Finalization
No finalization steps.