• Topic ID: id_18718287
  • Version: 2.0
  • Date: Feb 4, 2020 1:40:31 AM

Gantry Cover Touch Pad Replacement

Prerequisites

Overview

Procedure Effectivity:

Procedure

  1. Remove side, top, and front covers.
  2. Notice Image
    • notice
    • Always turn OFF the HVDC before the 120 VAC. Turning OFF 120 VAC power before HVDC power can result in equipment damage.
  3. Turn off the three (3) main power switches (Axial Drive, HVDC, 120VAC) on the Service Switch Panel.
  4. Disconnect touch strip cable.
  5. Remove 2 nuts that secure touch pad to outside of gantry.
  6. Install new touch pad.
  7. Notice Image
    • notice
    • Always turn ON the 120 VAC before the HVDC. Turning ON HVDC power before 120 VAC power can result in equipment damage.
  8. Reassemble gantry.

Finalization

No finalization steps.