Filament Calibration
Prerequisites
Run the Filament Calibration procedure. The Filament Calibration tool runs an automated scan sequence and populates the JEDI database accordingly.
Procedure
- Launch the GENERATOR TOOL from the Service Desktop. See Software section for more details.
- Launch the Filament Calibration tool.
- Run the test.
- The software automatically warms up the tube and run the scans to calibrate
the filament. If filament calibration fails, run the program again.note:
If Filament Calibration fails persistently, try downloading the default database to the PPC kV control board and running this procedure again.
Finalization
No finalization steps.