- Topic ID: id_18718132
- Version: 2.0
- Date: Feb 4, 2020 1:40:31 AM
Control/Scan Start Panel Replacement
Prerequisites
Overview
Procedure
- Remove gantry covers as required.
Refer to
- notice
- Turn off the three power switches (Axial Drive, HVDC, 120VAC) on the Service Switch Panel.
- Disconnect J1 and left control cables.
- Loosen the five screws (For BrightSpeed, four screws) that fasten the start panel to the cover.
- Replace the control panel and reconnect J1 and left control cables.
- notice
- Reassemble the gantry.
![]() |
![]() |
Finalization
- Perform a functional scan to confirm proper operation.