- Topic ID: id_23554302
- Version: 1.0
- Date: Oct 9, 2018 1:36:44 PM
CTPM-1708 Cradle bottom inspection (For High Capacity Table)
Prerequisites
Overview
Procedure Effectivity:
Procedure
- Raise the Table to maximum height.
- Move the Cradle and IMS to OUT limit position.
- Remove power from Table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
- Check the bottom of the cradle. There should be no big scratch.
- Wipe the bottom of the cradle with alcohol.
- Wait a while for alcohol volatilization.
Finalization
- Power up the Table from the Service Switch Panel.
- Verify that the cradle and IMS movement are smooth and normal.