48cm Phantom Image Series Verification
Prerequisites
1 Acquiring the 48cm Phantom Image Series
Procedure
- Mount the Phantom Holder on the head-end of the table.
- Mount the 48cm Phantom on the Phantom Holder.
- Align, level, and center the 48cm Phantom.
-
Align phantom using the internal laser lights.
-
Level phantom using bubble level and the Z Axis knob on the Phantom Holder.
-
Center phantom using the CENTER PHANTOM utility in the left head SCANNER UTILITIES selection and the X and Y Axis knobs on the Phantom Holder.
-
- Set up the system to scan a four scan, 16 image, 48cm Phantom image
series.
MANUAL SCAN PROTOCOL SETUP
Refer to Table 2 and setup an Axial scan with the parameters as shown.
AUTO SCAN PROTOCOL SETUP
-
On the Exam Rx desktop, select NEW PATIENT.
-
Type the following entries in the two listed Patient Information fields:
Patient ID: Service
Name: 48cm Phantom Image Series
-
From the Protocol display, click on the infant box.
-
On the infant display window, click on the area below the infant’s right foot to display the Miscellaneous menu.
-
Click on the 20:10 IMAGE SERIES 48 CM button.
-
Select the first series (Series Description filed should display 48cm Lg Series 1)
-
- Set internal Landmark.
- Acquire a four-scan, 16 image, 48 cm Phantom image series.
2 Brightness Uniformity and Noise
Procedure
- Specifications:
Each Row (2A, 1A, 1B, and 2B) of the series must pass 48cm Brightness Uniformity and Noise (for the first series scan parameters) specifications:
- Recommended Failure Recovery:
- Perform DETAILED CAL.
- Perform AUTO CT# ADJUST.
- Repeat this procedure to verify 48cm Phantom Image Performance.
3 Finalization
Finalization
No finalization steps.