35cm Phantom Image Series Verification (RT16)
Prerequisites
Overview
1 Acquiring the 35cm Phantom Image Series
Procedure
- Mount the Phantom Holder on the head-end of the table.
- Mount the 35cm Phantom on the Phantom holder.
- Align, level, and center the 35cm Phantom.
-
Align phantom using the internal laser lights.
-
Level phantom using bubble level and the Z Axis knob on the Phantom Holder.
-
Center phantom using the CENTER PHANTOM procedure in the left head SCANNER UTILITIES selection and the X and Y Axis knobs on the Phantom Holder.
-
- Set up the system to scan a single, 64 image or 16 image, 35cm
Phantom series.
MANUAL SCAN PROTOCOL SETUP
Refer to Table 2 and setup an Axial scan with the parameters shown.
AUTO SCAN PROTOCOL SETUP
- Acquire a single, image series of the 35 cm Phantom.
2 Image Performance Verification
2.1 Series Means (AvXo-AvXc and AvSDo) & Center Smudge
Procedure
- Launch the Image Analysis Tool.
- From the Global Control Palette, click on the Service Desktop.
- From the Service Desktop upper navigational bar, click on the IMAGE QUALITY icon.
- In the left navigational area directory, click on IMAGE ANALYSIS.note:
During the Image Analysis process, you will need to switch between the Service Browser screen (to select the image series to analyze) and the Image Analysis screen (to run a particular Auto 1x analysis test). Use ALT/F1 to bring a window to the front. Use ALT/F3 to send a window to the back.
- Using the AUTO 1X Tool on the Image Analysis
screen, generate a Series Means and Center Smudge report for the 2nd
series of the 35cm Phantom acquisition.
- From the Service Browser window, select the 35cm Phantom series acquired in Acquiring the 35cm Phantom Image Series.
- On the Image Analysis Tool window, click on the AUTO
1X diamond shaped button.
- Click on the AUTO 1X rectangular button, select the IMGSER 35CM test from the pull-down menu, slide cursor to the right and select 16X1.25/140KV/550MA/1S or 4X3.75/120KV/200MA/3S from the test pull-down menu, and continue to slide cursor to the right to select AUTO.
- Click on the ACCEPT button.
- Review the Image Analysis Report data and record the Series Means (AvXo-AvXc, and AvSDc) data and the Center Smudge data for each row in Table 3.
2.2 Band and Streak Artifact
Procedure
- Review the 35cm Phantom images acquired in Acquiring the 35cm Phantom Image Series for Bands and
Streaks using the Image Works Desktop.
- From the Global Control Palette, click on the IMAGE WORKS Desktop.
- From the Image Works Browser window, select the 35cm Phantom series acquired in Acquiring the 35cm Phantom Image Series.
- From the Image Works Browser window, click on the VIEWER button.
- While paging through the 64 images in the series, analyze images for any Band or Streak artifacts. Adjust Window/Level using the center mouse button, as required.
- Note any image that appears to have a Band or Streak artifact.
- Verify any images noted above (Step 1.e) meet 35cm Phantom
Band or Streak Image Performance requirements.
- From the Global Control Palette, click on the Service Desktop.
- From the Service Browser window, select an image of the 35cm Phantom series acquired in Acquiring the 35cm Phantom Image Series that appeared to have a Band or Streak while reviewing the series in Image Works.
- On the Image Analysis Tool window, click on the AUTO 1X rectangular button, select the IMGSER 35CM test from the pull-down menu, slide cursor to the right and select 16X1.25/140KV/550MA/1S or 4X3.75/120KV/200MA/3S from the test pull-down menu, and continue to slide cursor to the right to select BAND or STREAK.
- Click on the ACCEPT button.note:
If required, adjust Window/Level using the center mouse button.
- Position and size the Band or Streak ROI using the left cursor
button.note:
To ensure specifications are properly applied, do not adjust the Reference ROI's for the Band and Streak artifact tests. The Reference ROI's are adjustable after the 1st Accept Modification click.
- Click on ACCEPT MODIFICATION twice, to generate a report.
- Verify the Overall Test Pass/Fail Indicator Window of the Report display indicates PASS.
- Repeat steps b through g, for each image noted in Step 1e.
Box Size = 173 mm2 to 408 mm2 17 mm (+/- 3.5 mm) x 17 mm (+/- 3.5 mm) 19 (+/- 4 pixels) x 19 (+/- 4 pixels)
Box Positions: Box 1 = 0 mm x 0 mm Box 2 = 0 mm x -128 mm Box 3 = 128 mm x 0 mm Box 4 = 0 mm x 128 mm Box 5 = -128 mm x 0 mm
3 Failure Recommended Actions
3.1 Series Means (AvXo-AvXc or AvSDo) Failure Recovery
Procedure
- Specifications:
Each Row of the series must pass 35cm Series Means (for the first series scan parameters) specifications:
-
AvXo - AvXc: < +/- 3.
-
AvSDo: < 17.0 for 16 slice; AvSDo: < 10.0 for 8 slice.
-
- Recommended Recovery:
-
Repeat Detailed Cal for the 35cm Phantom.
-
Repeat Acquiring the 35cm Phantom Image Series and Image Performance Verification to verify Image Performance.
-
3.2 Center Smudge Failure Recovery
Procedure
- Specifications:
80% the Rows must pass Center Smudge specifications when re-scanning:
Center Smudge: Smudge factor < +/- 4.0
- Recommended Recovery:
- If one image of one row fails (row fails once):
-
Re-scan the 35cm Phantom acquiring two series of images using the same protocol settings as listed in Table 2).
-
Analyze the two 64 re-scanned image series using the Image Analysis IMGSER 35CM -> 16X1.25/140KV/550MA/1s or 4X3.75/120KV/200MA/3s -> AUTO tool.
-
Verify that all the 128 re-scanned images do not fail Series Means (AvXo - AvXc and AvSDo) or Center Smudge.
-
- If more than one image of one Row fails (Row fails more than
once) or multiple rows are failing:
-
Repeat Detailed Cal for the 35cm Phantom.
-
Re-scan the 35cm Phantom acquiring three series of 64 images using the same protocol settings as listed in Table 2.
-
Analyze the three re-scanned image series using the Image Analysis IMGSER 35CM -> 16X1.25/140KV/550MA or 4X3.75/120KV/200MA/3s -> AUTO tool.
-
Verify that all the re-scanned images do not fail Series Means (AvXo - AvXc and AvSDo) or Center Smudge.
-
- If one image of one row fails (row fails once):
3.3 Band or Streak Artifact Failure Recovery
Procedure
- Specifications:
-
Band Artifact: Band Factor < +/- 2.5 where radius=0~14cm;
Band Artifact: Band Factor < +/- 3.5 where radius=14~15cm;
Band Artifact: Band Factor < +/- 12 where radius>15cm
-
Streak Artifact: Streak Factor < +/- 4.0.
-
- Recommended Recovery:
- Troubleshoot Band or Streak artifact failures (refer to “Troubleshooting” category).
- Repeat Detailed Cal for the 35cm Phantom.
- Repeat Acquiring the 35cm Phantom Image Series and Image Performance Verification to verify Image Performance.