• Topic ID: id_15460419
  • Version: 3.0
  • Date: Jun 15, 2020 10:58:22 PM

Generator Data (Pro, RT, VCT)

This is the list of the Jedi HP X-Ray Generator tracked data. It includes all the critical component temperatures measured or calculated with the generator thermal model (basing on the measurement of the inverter heatsink surface), the DC bus measurement and the cumulated exposure time.

Max Temperatures for Heatsink

Max Temperatures for Parallel Inductor

Max Temperatures for MPHV Tank

DC_bus min in Preparation

DC_bus max in Preparation

DC_bus min during exposure

DC_bus max during exposure

Time since first power-up

NVRAM battery life time

Cumulated Power-on time

Inverter average current

Actual kV

Actual mA

Actual exposure time