Streak Test
Prerequisites
Overview
This document provides the necessary steps to test for streak artifacts.
1 Phantom Setup
Procedure
- Move table to longitudinal home position.
- Attach phantom holder to gantry end of cradle.
- Mount phantom to phantom holder.
- Place bubble level on top of phantom, positioned along Z-axis. Adjust phantom holder back knob until phantom is level front to back.
- Turn on laser alignment lights.
- Move table into gantry until axial internal alignment lights are aligned with phantom's axial reference mark.
- Click SCANNER UTILITIES.
- Click CENTER PHANTOM.
- Click CONFIRM.
- Press START on SCIM when it flashes.
- Adjust phantom position if results are out of specification
range.note:
Phantom center specification is ± 0.5mm.
- Repeat steps [9] – [11] until phantom is centered.
- Click DONE in Phantom Centering window.
- Click QUIT in Scanner Utilities window.
- Set Internal Landmark.
2 Acquire Scans
Procedure
- Click NEW PATIENT.
- Type the following entries in the listed Patient Information fields:
- Patient ID: Service
- Patient Name: Streak Test
- Select Service tab from Protocol Selection, Anatomical Selector.
- Select IMAGE QUALITY, 44.5
STREAK TEST.note:
Make sure the prescription is a single group axial protocol.
- Click CONFIRM.
- Click CONTINUE in Attention window.
- Press START on SCIM when it flashes.
- Click END EXAM.
- If PPS Response window is displayed, click COMPLETE.
3 Image Review
Procedure
- Select ImageWorks desktop.
- Select scanned examination from List/Select browser.
- Click VIEWER and review all images for streak artifacts.
- Perform Streak Analysis on images with streak artifacts.
4 Streak Analysis
Procedure
- Select SERVICE desktop and move Service desktop to left monitor.
- Select IMAGE QUALITY from Service Desktop.
- Select IMAGE ANALYSIS and move Image Analysis window to left monitor.
- Select image from right monitor List/Select browser.
- Click MANUAL radio button on Image Analysis window.
- Click MANUAL selection button, STREAKS.
- Click ACCEPT. The image is displayed in the Image Analysis window with a Streak ROI.
- Follow instructions in Image Analysis window lower right pane to size Streak ROI and analyze.
- From Report window, if Measured Streak Factor is out of specification range +/-4.0, perform Image Quality Troubleshooting in accordance with prescribed procedures.
- Click DISMISS in Report window.
- Click QUIT in Image Analysis window.
5 Finalization
No finalization steps.