• Topic ID: id_18480126
  • Version: 4.0
  • Date: Jan 20, 2020 8:32:43 PM

NIO64 Host Computer (Z820) FDIP Card Replacement

Prerequisites

Overview

This procedure shall be followed when replacing the FDIP Card in the NIO64 host computer (Z820).

1 Z820 Computer Removal

Procedure

  1. Open the Z820 computer side access panel.
  2. Remove the Expansion Card Support.

    Figure 1. Z820 Airflow Guide and Expansion Card Support

  3. Replace the existing FDIP card with new one.

    Figure 2. Z820 Component Location

  4. Install the Expansion Card Support and close the Side Access Panel.

2 Restore the Console

Procedure

  1. Fully slide the Z820 computer back into the console chassis, until mounting brackets are flush against the console chassis.
  2. Install and torque the four (4) 3mm socket cap screws that mount the Z820 computer in the NIO64 console.
    note:

    Torque each the for (4) 3mm socket cap mounting screws to:

  3. Reconnect all cables removed earlier to the Z820 computer.
  4. Remove LOTO on console.

3 Finalization

Procedure

  1. Install console covers. Refer to Replacement → Console (RIO) → Console Cover Removal and Installation Procedure.
  2. Perform the Functional Checks → System Scanning Test instructions from the procedure list.