- Topic ID: id_11038857
- Version: 3.0
- Date: Dec 3, 2019 2:00:48 AM
HP Host Computer Z800 System/Image HDD Replacement
Prerequisites
Overview
Procedure
- Select one of the following methods to Power OFF the Operator
Console:
- If Applications are up, click on the Shut Down button and select Shutdown.
The Operator Console monitor will display a 'Power Down' message when it is acceptable to power OFF the Operator Console.
- If Applications are down, open a Unix Shell. Type: halt.
- If Applications are up, click on the Shut Down button and select Shutdown.
- Power OFF the console at the front panel switch.
- Remove the front and left side covers from the console.
Refer to
- Remove the following components from the host computer (Z800):
-
Side Access Panel
-
Airflow Guide
-
Expansion Card Support
Figure 1. Airflow Guide and Expansion Card Support

-
- Disconnect the power and signal cables from two system hard
disk drives (HDDs).
Figure 2. Two System Hard Disk Drive

- Pull out the hard disk drive (HDD) tray from front side.
Figure 3. Drive Release Latch

- Replace hard disk drive (HDD) with new one.
- Reinstall all removed components, and reconnect any cables that have been disconnected.
- Refer to LFC procedure to load all the software and options.
Finalization
- Refer to System Scanning Test to confirm proper operation.
- Reinstall the console covers.