Fast Calibration
Prerequisites
Overview
Fast Calibration ensures consistent image quality and radiation exposure by correcting for channel gain differences in the detector, detector drift correction resulting from radiation damage, imperfection in the bowtie filter, and removal of bowtie shape from the data.
Procedure
- Move table to longitudinal home position.
- Click DAILY PREP.
- Click FAST CALIBRATION.
- Press START on SCIM/GSCB when it flashes.
- Click QUIT in Daily Preparation window.
Finalization
note: If this procedure has been done as part of Tube Replacement procedure, return to Tube replacement procedure for remaining steps.