Cradle Lock Assy Replacement
Prerequisites
Overview
Procedure
- Raise the Table to maximum height.
- Move the Cradle and IMS to OUT limit position.
- Remove power from Table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
- Remove the following Table covers:
-
Top Cover (Right/Left)
-
Rear Top Cover
-
Rail Cover (Right)
-
- Disconnect the Lock Assy cable connector.
- Unscrew 3 screws, and remove the Lock Assy from the Table.
Figure 1. Cradle Lock Assy Removal

- Install the new Lock Assy into place, and tighten the 3 screws.
- Push the cylinder head to release the cradle lock, and verify
that the gap between the lock block and the cradle lock is 1.5mm ±
0.25.
If not, adjust the lock block position, or/and adjust the cradle lock position using adjusting screw.
Figure 2. Cradle Lock Adjustment

- Release the cylinder head, and verify that the cradle locking function works well.
- Connect the Lock Assy cable connector.
Finalization
- Re-install the Table covers.
- Power up the Table from the Service Switch Panel.
- Verify that the cradle IN/OUT function is operating normally.