- Topic ID: id_18480159
- Version: 3.0
- Date: Jan 20, 2020 8:35:16 PM
(12HW28.8) Service Pack 3 Installation Procedure
Prerequisites
Overview
This procedure describes 12HW28.8 Service Pack 3.0 installation procedure.
The Service Pack 3.0 software CDROM contains Service Pack 1.0, 2.0 and 3.0 patches.
1 Check software version currently installed
Procedure
- Open a Unix shell.
- Type swhwinfo ENTER
Confirm that the base software version is 12HW28.8.
2 Service Pack Installation
Procedure
- Perform Save System State (select [All] button).
- If Exam Split option is installed, record the mode (HES or VES)
by the following command.
- Open Unix shell and and log on as root.
- Type ls -l ~ctuser/ves/.hesMode
- If there is .hesMode file, it is HES mode. If not, It is VES mode
- Uninstall the following options if they are installed.
-
AutoBone_Xpress
-
AVA_Xpress
-
CardEP
-
CardIQ_Xpress_Reveal
-
CT_Colono_Pro3D_EC
-
CT_Perfusion_4D_MultiOrgan
-
CT_Perfusion_4D_Neuro
-
Volume Viewer
-
Exam Split
-
- Open a Unix shell.
- If Application software is up, type the following command.
{ctuser@hostname} cleanMon
- In the Unix Shell, become root:
- {ctuser@hostname} su -
- Password: <password>
- Insert the Service Pack CD into the Host DVD drive
- Perform the following steps to install the Service Pack CD:
-
[root@hostname]#
patch_install
-c
note:
If Service Pack 1.0 is already installed, go to step e.
note:If Service Pack 2.0 is already installed, go to step h.
- The following message is displayed
These patches from cdrom will be installed:
12HW28.8_GEHCServicePack1.0
12HW28.8_GEHCServicePack2.0
12HW28.8_GEHCServicePack3.0
-------------------------------------------
I will install update 12HW28.8_GEHCServicePack1.0 from CDROM, is this ok ? [y/n]
- Input y to install Service Pack 1.0.
- Confirm the “Update 12HW28.8_GEHCServicePack1.0 has been installed successfully” message.
- The following message is displayed.
I will install update 12HW28.8_GEHCServicePack2.0 from CDROM, is this ok ? [y/n]
- Input y to install Service Pack 2.0.
- Confirm the “Update 12HW28.8_GEHCServicePack2.0 has been installed successfully” message.
- The following message is displayed.
I will install update 12HW28.8_GEHCServicePack3.0 from CDROM, is this ok ? [y/n]
- Input y to install Service Pack 3.0.
- Confirm the “Update 12HW28.8_GEHCServicePack3.0 has been installed successfully” message.
-
[root@hostname]#
patch_install
-c
- Check Software Version:
[root@hostname]# swhwinfo ENTER
- System should reply with a response similar to the following:
12HW28.8_SP3-0-x. <hw info (ex. H40_P2_SS64_G_GTL) >
- Remove the Service Pack CD from Host DVD drive.
- Reboot the system:
[root@hostname]# reboot
- Perform Restore System State. (Select correct HES/VES setting if asked by system.)
- Restart the system:
- Confirm all option keys are restored correctly by Verify Options in Common Service Desktop.
3 How to uninstall Service Pack
Procedure
- Perform Save System State (select [All] button)
- If Exam Split option is installed, record the mode (HES or VES)
by the following command.
- Open Unix shell and and log on as root.
- Type ls -l ~ctuser/ves/.hesMode
- If there is .hesMode file, it is HES mode. If not, It is VES mode.
- Uninstall the following options if they are installed.
-
AutoBone_Xpress
-
AVA_Xpress
-
CardEP
-
CardIQ_Xpress_Reveal
-
CT_Colono_Pro3D_EC
-
CT_Perfusion_4D_MultiOrgan
-
CT_Perfusion_4D_Neuro
-
Volume Viewer
-
Exam Split
-
- Open a Unix shell.
- If application software is up, type the following command.
{ctuser@hostname} cleanMon
- In the Unix Shell, become root:
- {ctuser@hostname} su -
- Password: <password>
- Perform the following steps to uninstall the Service Pack CD:
- [root@hostname]# patch_install –u
- The following message is displayed.
The following patches are installed on system:
----------------------------------------------
12HW28.8_GEHCServicePack3.0
12HW28.8_GEHCServicePack2.0
12HW28.8_GEHCServicePack1.0
----------------------------------------------
I will uninstall update 12HW28.8_GEHCServicePack3.0, is this ok? [y/n]
- Input y to uninstall Service Pack 3.0.
- The following message is displayed.
I will uninstall update 12HW28.8_GEHCServicePack2.0, is this ok? [y/n]
- Input y to uninstall Service Pack 2.0.
- The following message is displayed.
I will uninstall update 12HW28.8_GEHCServicePack1.0, is this ok? [y/n]
- Input y to uninstall Service Pack 1.0.
- Reboot the system:
[root@hostname]# reboot
- Restore System State. (Select appropriate HES/VES setting if asked by system.)
- Restart the system:
- Confirm all option keys are restored correctly by Verify Options in Common Service Desktop.
4 Finalization
No finalization steps.