(10HW33.x) Service Pack Installation Procedure
Prerequisites
Overview
Procedure
- If Application software is up, perform the following command
to shutdown the application.
{ctuser@hostname} cleanMon ENTER
- In the Unix Shell, become root:
- {ctuser@hostname} su – ENTER
- Password: #bigguy ENTER
- Insert the Service Pack CD into DVD drive in the host.
- Perform the following steps to install the Service Pack CD:
- [root@hostname]# patch_install –c
- Input “y” to install the patch.
- After the patch installation process completes, verify that
the Service Pack loaded correctly:
- {root@hostname} swhwinfo
- Confirm right software version.
- Remove the Service Pack CD from Host DVDROM drive.
- Type : [root@hostname] # sync
[root@hostname] # sync
[root@hostname] # reboot
- Perform the Flash Download (if required).
- Restore system state.note:
When install Class M software in the system which a Service Pack (SP2.0 or SP3.0) is installed in, please to in the following procedures:
-
Uninstall SP2.0 and SP3.0 software.
-
Install Class M software.
-
Install SP2.0 and SP3.0 software again.
Service Pack (SP2.0) that included in Service Pack (SP3) Media contains the update of the Class M software. (If Class M software is installed after SP2.0 installation, the update contents of the Class M software in SP2.0 is missing.)
note:This Service Pack (SP2.0) updates the JEDI database. Therefore it is necessary to perform Filament Cal after Flash Download.
note:This procedure can be done at OC line but need to perform Filament Cal after Flash Download at the system.
-
Finalization
No finalization steps.