Rotor Diagnostic Test (LS16, Ultra, Plus)

Overview

Objective: The purpose of this test is to drive the rotation inverter in high speed mode and low speed mode on the tube in order to identify a defective rotation board or defective phase capacitors or a wrong connection among rotation board, HEMIT and tube. Summary: Run diagnostic. The test is executed without high voltage and filament heating, and needs no manual operation.

  1. Launch the GENERATOR TOOL.
  2. Launch the Rotor Diagnostic tool.
  3. Run the test. A rotor sequence will proceed.
  4. If a problem occurs, check the error code.

1 Finalization