• Topic ID: id_11038900
  • Version: 3.0
  • Date: Apr 22, 2019 12:56:16 AM

Heater Diagnostic Test (LS16, Ultra, Plus)

Overview

Objective: To drive the heater inverter on both tube filaments in order to identify a faulty heater board or a wrong connection between the heater board, HV Tank and tube. Summary: Run diagnostic. The test is executed without high voltage or anode rotation, and needs no manual interaction.

  1. Launch the GENERATOR TOOL.
  2. Launch the HEATER DIAGNOSTIC tool.
  3. Run the test. A scan sequence will proceed.
  4. If a problem occurs, check the error code.

1 Finalization

No finalization steps.