| Optima CT660 Service Methods (General), Direction 5366080-8EN |
| Rev |
Date |
Reason For Change |
| 40 |
May 11, 2022 |
Replacement > Table: Added "Table Speaket Replacement for GT/Lite Table" |
| 39 |
February 22, 2022 |
Added Smart Subscription Section in the manual
Install > Smart Subscription > EHL1.5: Added "Smart Subscription Connect", added "ASiR-V Feature Connection", added "SmartMAR Feature Connection", added "AW Server Feature Connection", added "SnapShot Freeze 2.0 Feature Connection".
|
|
Replacement > GT2000 / GT2000x Table : Added "Cradle Drive Rolly Assy Replacement".
Replacement > GT1700V Table : Added "Cradle Drive Rolly Assy Replacement".
|
| 38 |
January 20, 2022 |
Install > Option : Added "Remote Control Kit with Assisted Video Monitoring System Service Manual (5863844-1EN)", added "Image Protocol Manager 1.0 Installation".
|
|
Software > Software Installation Procedure (LFC) : Added "15HW25.2 Service Pack 3.3 Installation Procedure", Added "16HW45.1 Service Pack 1.3 Installation Procedure", Added "20HW16.10 Service Pack 1.0 Installation Procedure".
|
|
Replacement > Gantry > High Voltage : Added "HV Cable Replacement".
|
| Planned Maintenance and Tasks > Warranty PM (1 Time/Year During Warranty Period): Updated "Planned Maintenance Schedule W" |
| Planned Maintenance and Tasks > Out of Warranty PM (3 Times/Year Alternating Schedules): Updated "Planned Maintenance Schedule A", Updated "Planned Maintenance Schedule B", Updated "Planned Maintenance Schedule C". |
| 37 |
January 14, 2021 |
Safery : Updated "Torque Wrench Information".
|
|
Install > Option : Added "Anzai Respiratory Gating System Installation", updated "Cardiac Monitor (IVY) Installation".
|
|
Software : Updated "Connecting Laptop to Console (16HW14.x or Later)".
|
|
Software > Software Installation Procedure (LFC) : Added "(16HW45.1) Service Pack 1.2 Installation Procedure", Updated "(20HW16.x) Load From Cold".
|
|
System Diagrams : Updated "Optima CT660 Functional Interconnects".
|
|
Part List : Updated "Gantry Parts List".
|
|
Replacement > Gantry > Tube : Updated "X-Ray Tube Replacement".
|
|
Replacement > Table > GT2000/GT2000x Table : Added "Elevation Potentiometer Replacement".
|
|
Replacement > Table > GT1700V Table : Added "Elevation Potentiometer Replacement".
|
| 36 |
November 06, 2020 |
Align, Setup, Cals > Table > Lite Table : Added Z8G4 Host PC BIOS Setup and CMOS Battery Replacement.
|
|
Functional Checks > Gantry : Updated Collimator Functional Diagnostic Tests.
|
| Software : Added Configure Product Network Filters (18HW46.x or later), Connecting Laptop to Console (16HW14.x or Later), Host Computer Image Database Reset (19HW19.x), Host Computer Image Database reset (20HW16.x or later), Updated Host Computer Image Database Reset. |
| Software > Software Installation Procedure (LFC) : Added (20HW16.x) Load From Cold, (20HW16.x) LFC Quick Guide. |
|
Theory > Console : Added Open Console (Z8G4) Theory.
|
|
System Diagrams : Updated Optima CT660 Functional Interconnects.
|
|
Troubleshooting > Console : Added OpenOC (Z8G4) Computer Troubleshooting, Image Database troubleshooting (20HW16.x).
|
|
Replacement > Console (RIO) : Addd Optima CT660/Revolution EVO Console Hardware Replacement Kit (Z8G4) Installation Manual (5858809-1EN).
|
|
Replacement > Console (NIO64) : Addd Additional procedure for NIO64 with Z8G4, Optima CT660/Revolution EVO Console Hardware Replacement Kit (Z8G4) Installation Manual (5858809-1EN).
|
|
Replacement : Addd Console (RIO/NIO64 with Z8G4).
|
|
Replacement > Gantry > Axial : Updated Axial Encoder Assembly Replacement.
|
|
Replacement > Gantry > Tube : Updated X-Ray Tube Replacement.
|
| 35 |
May 22, 2020 |
Installation > Option : Added Optima CT660 Phase 2.5 IB Upgrade Kit Installation Manual (Dir 5599284-1EN).
|
|
Align, Setup, Calibrations > Gantry > Tube: Updated HOT ISO Alignment.
|
|
Align, Setup, Cals > Table > Lite Table : Added Cradle Belt Tension Adjustment
|
|
Software > Software Installation Procedure (LFC) : Added (16HW45.1) Service Pack 1.1 Installation Procedure.
|
|
Troubleshooting > Console : Added "Z820/Z840 RAID Analysis and Configuration".
|
|
Part List : Updated Gantry Parts List, Table Parts List, PDU Parts List.
|
|
Troubleshooting > Table : Added GT Trouble Shooting.
|
|
Troubleshooting > X-Ray Generation : Updated HEMIT Air Removal Procedure.
|
|
Replacement > Gantry > DAS (SDCB) : Added Rotating 48V Power Supply Replacement
|
|
Replacement > Gantry > DAS (HALO) : Added Rotating 48V Power Supply Replacement
|
|
Replacement > Table > GT2000/GT2000x Table : Added GT Top Side Cover Bracket Replacement, GTCB Cover Replacement.
|
|
Replacement > Table > GT1700V Table : Added GT Top Side Cover Bracket Replacement, GTCB Cover Replacement.
|
|
Replacement : Addd PDU.
|
|
Planned Maintenance and Tasks > 3 Times PM / Year: Updated Planned Maintenance Schedule A, Planned Maintenance Schedule B, Planned Maintenance Schedule C, Planned Maintenance Schedule W.
|
| 34 |
January 21, 2020 |
Installation : Updated “Optima CT660 Pre-Installation”.
|
|
Software > Software Installation Procedure (LFC) : Updated (10HW02.x / 10HW33.x) Load From Cold - Optima CT660, (11HW34.x / 12HW12.x) Load From Cold - Optima CT660, (12HW28.x) Load From Cold - Optima CT660, (13HW38.x) Load From Cold - Optima CT660, (14HW30.x) Load From Cold - Optima CT660, (15HW13.x) Load From Cold, (15HW25.x) Load From Cold - Optima CT660, (16HW45.x) Load From Cold.
|
|
Replacement > Gantry > Tube : Updated “X-Ray Tube Replacement”.
|
| 33 |
December 3, 2019 |
Software > Software Installation Procedure (LFC) : Updated (10HW02.x / 10HW33.x) Load From Cold - Optima CT660, (11HW34.x / 12HW12.x) Load From Cold - Optima CT660, (12HW28.x) Load From Cold - Optima CT660, (13HW38.x) Load From Cold - Optima CT660, (14HW30.x) Load From Cold - Optima CT660, (15HW13.x) Load From Cold, (15HW25.x) Load From Cold - Optima CT660, (16HW45.x) Load From Cold.
|
| 32 |
September 26, 2018 |
Safety : Updated “Equipment Service - Lockout-Tagout-PPE”. |
| Installation Updated “Optima CT660 Pre-Installation”. |
Replacement > Gantry :- > High Voltage : Updated “HEMIT Tank Replacement”, “Power Unit Replacement”.
- > Tube : Updated “X-Ray Tube Replacement”.
|
| Planned Maintenance and Tasks : Added “HPM Overview”, “Warranty PM (1 Time/Year During Warranty Period)”, “Out of Warranty PM (3 Times/Year Alternating Schedules)”. |
| 31 |
May 21, 2018 |
Installation > Option : Added “Table Protection Pad Installation (Dir 5786404-1EN)” |
| Software > Software Installation Procedures (LFC) : Added “(15HW25.2) Service Pack 3.1 Installation Procedure”, “(16HW45.1) Service Pack 2.0 Installation Procedure”, “Install Software Options by eLicense”. |
| Replacement > Gantry : Updated “CDCB Replacement - Merc40” |
| 30 |
June 08, 2017 |
Installation > Option : Added “Optima CT660 / Revvolution EVO VT2000 Table Replacement” |
Software :- Added “Audit Trail”. Updated “SSA License File Installation”
- > Software Installation Procedures (LFC) : Added “(15HW25.2) Service Pack 2.0 Installation Procedure”.
|
| Part List : Updated “Console Parts List” |
Replacement :- > Console (RIO) : Updated “LCD Monitor Replacement”
- > Console (NIO64) : Updated “LCD Monitor Replacement”
- > Gantry > DAS and Detector Common : Updated “FRDM Replacement Procedure”
|
| 29 |
March 22, 2017 |
Align, Setup, Calibrations > Console : Updated “LCD Video Monitor Setup for Z800” |
| Functional Checks > Console : Updated “Peripheral Tower Functional Test (PMT-20/PMT-21/PMT-22)” |
| Software > Software Installation Procedures (LFC) : Added “(16HW45.1) Service Pack 1.0 Installation Procedure” |
| Theory > Console : Updated “LCD Video Monitor Overview” |
| Troubleshooting > X-Ray Generation : Added “HEMIT Air Removal Procedure” |
Replacement :- > Console (RIO) : Updated “HP PC Battery Replacement Procedure”, “LCD Monitor Replacement”
- > Console (NIO64) : Updated “HP PC Battery Replacement Procedure”, “LCD Monitor Replacement”
- > Gantry > Tube : Updated “X-Ray Tube Replacement”
|
| 28 |
December 21, 2016 |
Readme : Updated “Optima CT660 Service Software Tools Classification ” |
Installation :- Updated “Pre-Installation”, “Installation Forms”
- > Option : Added “Optima CT660 / Revolution EVO MAR Upgrade Procedure”. Updated “AWE Connection Option Installation”. Removed “Liquid Bearing Tube IB Two-Step Upgrade”
|
Align, Setup, Calibrations :- > Console : Added “How to add Body-Part Examined Tag (0018,0015) to Image Headers”
- > Gantry : Updated “Configuring Gantry Tilt for Extender Removal”
|
Software :- Updated “SSA License File Installation”
- > Software Installation Procedures (LFC) : Added “(16HW45.x) Load From Cold”, “(16HW45.x) LFC Quick Guide”.
|
| Part List : Updated “Gantry Parts List”, “Table Parts List” |
Replacement > Gantry :- > DAS (Merc40) : Updated “Rotating 48VDC 1kW Power Supply Replacement“, “Rotating 48VDC 1kW Power Supply Replacement from Original 1.8kW Power Supply“
- > Display : Updated “Xtream Display CF Memory Card Replacement”
- > ORP and TGP : Updated “ORP Replacement (M40)“
|
Planned Maintenance and Tasks > 3 Times PM / Year :- Added “Planned Maintenance Schedule W”
- > PM Data Reports : Updated “Planned Maintenance Forms ALL Schedules”
|
| 27 |
April 15, 2016 |
Planned Maintenance and Tasks : Removed “4 Times PM / Year” |
| 26 |
February 08, 2013 |
Installation > Upgrade : Added “Liquid Bearing Tube IB Two-Step Upgrade ”, “Liquid Bearing Tube IB One-Time Upgrade”, “Ball Bearing Tube Restoration Procedure” |
| Software > Software Installation Procedures (LFC) : Updated “(15HW25.x) Load From Cold - Optima CT660”. Added “(15HW25.2) Service Pack 1.0 Installation Procedure” |
| 25 |
December 04, 2015 |
Installation : Added “Optima CT660 Liquid Bearing Tube IB Upgrade”- Updated “Pre-Installation”, “Installation Forms”
- > Option : Updated “Optima CT660 Liquid Bearing Tube IB Upgrade”, “Xtream Injector Option”. Added “Optima CT660 / Revolution EVO MAR Upgrade Kit Installation”. Removed “Optima CT660/Revolution EVO Phase3.5 IB Upgrade Kit Installation (Dir 5694258-1EN) “
|
Software :- Updated “SSA License File Installation”
- > Software Installation Procedures (LFC) : Added “(15HW13.7) Service Pack 1.0 Installation Procedure”
|
| Theory > Gantry > DAS and Detector : Updated “CDAS Theory” |
Troubleshooting :- > Console : Updated “Z800 Host PC Troubleshooting”. Added “Troubleshooting Information for Image Browser Blue line issue”
- > Gantry : Added “Merc40 Detector Known Issue”, “Merc40 Detector Power Distribution Troubleshooting”
- > Image Quality : Updated “Z alignment Cal. Troubleshooting”. Added “How to Setup Low Signal correction for SmartPrep”
- > X-Ray Generation : Added “Troubleshooting Information for Disabled Scan Resume Issue after Rotor Error”
|
| Part List : Updated “Gantry Parts List”, “Console Parts List” |
Replacement :- > Console (RIO) : Updated “HP Host Computer Z800 RAID Controller”. Added “HP PC Battery Replacement Procedure”
- > Console (NIO64) : Added “HP PC Battery Replacement Procedure”
- > Gantry > DAS and Detector Common : Updated “FRDM Replacement Procedure”. Added “Interposer Replacement Procedure”
- > Gantry > DAS (Merc40) : Updated “Detector Module Replacement - Merc40”
- > Table > GT2000/GT2000x Table : Added “Link Bar Assembly Replacement”
|
| Planned Maintenance and Tasks > 3 Times PM / Year : Updated “Planned Maintenance Schedule B” |
| 24 |
August 31, 2015 |
Installation > Option : Added “Optima CT660 Liquid Bearing Tube IB Upgrade” |
| Software > Software Installation Procedures (LFC) : Added “(15HW25.x) Load From Cold - Optima CT660” and “(15HW25.x) RIO LFC Quick Guide” |
| Part List : Updated “Gantry Parts List” |
| 23 |
August 07, 2015 |
Planned Maintenance and Tasks : Added “3 Times PM / Year” |
| 22 |
June 30, 2015 |
In addition to general procedure updates, the following major changes were made: |
| Functional Checks > Gantry : Updated “zalignmentWizard User Instructions (Merc40)” |
| Replacement > Gantry > DAS (Merc40) : Updated “Detector Module Replacement - Merc40” |
| 21 |
June 15, 2015 |
Installation :- Updated ”Pre-Installation”, “Installation Manual” and “Installation Forms
- > Option : Added “Cardiac Monitor (IVY) Installation”, “Optima CT660/Revolution EVO Phase3.5 IB Upgrade Kit Installation (Dir 5694258-1EN)”
|
Align, Setup, Calibrations :- > Console : Added “Data Privacy (EA3 User Authorization) Configuration (15HW13.x or later)” and “Dose Check Management and EA3 Configuration (15HW13.x or later) “. Updated “LCD Video Monitor Setup for Z800”
- > Gantry : Updated “Gantry Balance Procedure”, “Intercom Adjustment”, “Resetting the C-Pulse” and “Slip Ring Adjustments”
- > Gantry > Enclosure : Updated “Alignment Light Adjustment”. Added “Backlight brightness adjusting tool for Gantry Control Panel” and “Xtream Display Japanese Character Setting”
- > Gantry > Tilt : Updated “Gantry Tilt Pot Adjustment”
- > Gantry > Tube : Updated “Z-Alignment”
- > Table > GT2000 : Updated “Cradle Motor and Cradle Drive Belt Tension Adjustment”, “Elevation Potentiometer Belt Tension Adjustment” and “IMS Belt Tension Adjustment”. Added “Table Load Height Setting“
- > Table > GT1700V : Updated “Cradle Motor and Cradle Drive Belt Tension Adjustment” and “Elevation Potentiometer Belt Tension Adjustment”.
|
| Functional Checks > Gantry : Updated “Alignment Lights Visual Checks”, “Gantry Rotation Safety Check”, HV Tank Feedback Resistor Verification” and “Meter Verification (16, Ultra, Plus)” |
Software :- Updated “Unmount SSA key procedure”
- > Software Installation Procedures (LFC) : Added “(10HW02.10) Service Pack 1.0 Installation Procedure”, “(10HW33.8) Service Pack 1.0 Installation Procedure”, “(12HW28.8) Service Pack 1.0 Installation Procedure”, “(13HW38.14) Service Pack 1.0 Installation Procedure”, “(13HW38.14) Service Pack 2.0 Installation Procedure”, “(15HW13.x) Load From Cold” and “(15HW13.x) LFC Quick Guide”. Updated “(11HW34.x) Service Pack Installation Procedure” and “(13HW38.14) Service Pack 2.1 Installation Procedure”
|
| Theory > Console : Updated “LCD Video Monitor Overview” |
| System Diagrams : Updated “Optima CT660 Functional Interconnects ” |
Troubleshooting :- > Console : Updated “Real-In-On Console Diagnostics (10HW10.x / 10HW33.x / 11HW34.x)”, “NIO64 Console Diagnostics” and “Z820 Host PC Troubleshooting”.
- > Diagnostics : Added “DAS Tools mA Ratio test failure known issue (Merc40)”
- > Gantry : Updated “Collimator Cleaning Procedure”
- > Image Quality : Added “Z alignment Cal. Troubleshooting”
- > Table : Added “Precise T/G alignment”
- > X-Ray Generation : Updated “Shorted Inverter Test (LS16, Ultra, Plus)”
|
| Part List : Updated “Gantry Parts List”, “Console Parts List”, “NGPDU Part List”, “Table Parts List” and “Misc. Items Parts List” |
Replacement :- Updated “Fuse Replacement Procedures”. Added “NGPDU-81 ABB AF116 Kxg Contactor Replacement “
- > Console (RIO) : Updated “LCD Monitor Replacement”
- > Console (NIO64) : Updated “DVD Peripheral Tower Replacement (Single Bay)” and “HP Host Computer Z820 Replacement”
- > Gantry : Updated “Gantry Balance Sensor Hardware Replacement“, “Gantry Heater Fuse Holder Replacement” and “Simplified HVDC Power Pan ASM Replacement”
- > Gantry > Axial : Updated “Axial Drive Assembly Replacement“, “2010 Axial Drive Brake Resistor Replacement”, “2010 Axial Drive Controller Replacement”, “2010 Axial Drive KIM Board Replacement”, “2010 Axial Drive Power Supply Replacement”, “Axial Encoder Assembly Replacement”, “Belt Removal and Installation” and “Home Flag Sensor Board Assembly Replacement”
- > Gantry > Collimator : Updated “Cam Drive Motor and Flex Coupling Replacement”, “Collimator Assembly Replacement“, “Collimator Control Board (CCB) Replacement” and “Collimator Filter Removal and Replacement”
- > Gantry > DAS and Detector Common : Updated “DAS and Detector Air Filter Replacement”, “DAS Fan, Fan Plate Assembly Replacement”, “Detector Replacement Procedure”, “Detector Air Plenum Removal and Installation”, “Detector Thermistor Replacement”, “Detector Window Replacement”, “FRDM Replacement Procedure”, “Heater Controller and Fan Controller Replacement” and “Rotating 48VDC PS and Fuse Board Replacement”
- > Gantry > DAS (HALO) : Updated “Backplane Replacement Procedure”, “DCB and DIFB Replacement” and “DIFB Chassis EMC Strip Replacement”
- > Gantry > DAS (SDCB) : Updated “SDCB DAS FRU Replacement”, “SDCB Backplane Replacement” and “SDCB Fiber Optic Transceiver Replacement”
- > Gantry > DAS (Merc40) : Updated “Backplane Replacement - Merc40”, “CDCB Replacement - Merc40”, “CFHC Replacement “, “Detector Collimator Unit Replacement - Merc40”, “Detector Carbon Window Replacement - Merc40”, “Detector Heater Replacement (Top / Bottom)”, “Detector Module Replacement - Merc40”, “Detector Air Plenum Removal and Installation”, “Plenum Fan Replacement”, “Power Board Replacement - Merc40”, “Power Cable (CFHC ↔ Heater Routing Assembly and Fuse Board)”, “RC232 Communication Cable Replacement”,”Thermistor Assembly Replacement”,”Thermistor Replacement”. Added “Merc40 DAS-Detector Assembly Replacement Procedure“
- > Gantry > Display : Updated “Breathing Light Assembly Replacement”, “Gantry Cover Tilt Sensor Replacement”, “Operator Control Panel Replacement”, “Xtream Display Assembly Replacement”, “Xtream Display CF Memory Card Replacement”.
- > Gantry > Enclosure : Updated “Gantry Bore Cover Removal and Re-Install”, “Gantry Bore Cover Support Tool Usage”, “Gantry Cooling Fan Replacement”, “Gantry Front Cover Removal and Re-Install”, “Gantry Heater Replacement”, “Gantry Microphone with MIC Cover Replacement”, “Gantry Rear Cover Removal and Re-Install”, “Gantry Stationary (CFC) Fan Control Module Replacement”, “Gantry Top Covers Removal and Re-Install”, “Gantry Top Fan Replacement”, “Intercom Board Replacement”, “Laser Light Replacement, “Service outlet and E-Stop Button Replacement”.
- > Gantry > High Voltage : Updated “Auxiliary Assembly Replacement”, “Equipped FAN Inverter Replacement”, “HEMIT Tank Replacement”, “Inverter Mid Power Assembly Replacement”, “Power Unit Replacement”, “PI Filter Board Replacement (LS16, Ultra, Plus)”, “Programmed KV Control Board Replacement”.
- > Gantry > ORP and TGPU : Updated “ORP Replacement”, “ORP Replacement (M40)”, “Stationary 48 Volt Power Supply Replacement”, “TGPHD Replacement”, “TGPHD Power Supply Replacement”.
- > Gantry > Slipring : Updated “Dual Channel Fiber Rx Replacement”, “Dual Channel Tx Replacement”, “Slipring Brush Block Replacement”, “Slipring Power Supply Replacement”.
- > Gantry > Tilt : Updated “Tilt Potentiometer Replacement”, “Tilt Potentiometer Replacement (New)”, “Tilt Interference and Limit Switches Replacement”, “Hydraulic Tilt Motor Assembly Replacement”, “Gantry Tilt/Fan Relay (GTFR) Board Replacement”.
- > Gantry > Tube : Updated “X-Ray Tube Replacement”.
- > Table > GT2000 Table : Added “RIM IMS Cover Set Replacement “
- > Table > GT1700V Table : Added “RIM IMS Cover Set Replacement “
|
Planned Maintenance and Tasks :- Updated “Planned Maintenance Schedule D”
- > PM Data Reports : Updated “PM Report Schedule D”
|
| 20 |
January 08, 2015 |
In addition to general procedure updates, the following major changes were made: |
| Software > Software Installation Procedures (LFC) : Added “(13HW38.14) Service Pack 2.2 Installation Procedure” |
| Replacement > Gantry : Updated “Simplified HVDC Power Pan ASM Replacement” |
| 19 |
October 09, 2014 |
In addition to general procedure updates, the following major changes were made: |
Installation :- Updated “Installation Manual”
- > Option : Added “64 slice option Installation” and “RIO Console Long Cable Option Installation (Dir 5400875-1EN)”
|
| Align, Setup, Calibrations > Gantry : Updated “Gantry Balance Procedure” and “Intercom Adjustment“ |
| Functional Checks > Console : Updated “Peripheral Tower Functional Test (PMT-20) “ |
Software :- Added “SSA License File Installation” and “Unmount SSA key procedure”
- > Software Installation Procedures (LFC) : Added “(14HW30.x) Load From Cold - Optima CT660” and “(14HW30.x) RIO LFC Quick Guide”
|
| Theory Gantry > DAS and Detector : Updated “CDAS Theory” |
| Troubleshooting > Console : Updated “Command Line Troubleshooting - RIO” |
| Part List : Updated “Console Parts List” |
Replacement :- Updated ”Certified Component Replacement Retest Matrix” and “Certified Component Verification Form e4879”
- Gantry > DAS (Merc40) : Updated ”Detector Module Replacement - Merc40”
- Gantry > Enclosure : Updated ”Gantry Bore Cover Removal and Re-Install”, “Gantry Front Cover Removal and Re-Install”, “Gantry Side Covers Removal and Re-Install” and “Gantry Top Covers Removal and Re-Install”
- Table > GT2000 Table : Updated ”Removal / Installation of Table Covers”
- Table > GT1700 Table : Updated ”Removal / Installation of Table Covers (GT1700V)”
|
Planned Maintenance and Tasks :- Updated “Planned Maintenance Schedule A”, “Planned Maintenance Schedule B”, “Planned Maintenance Schedule C” and “Planned Maintenance Schedule D”
- > PM Data Reports : Updated “PM Report Schedule A”, “PM Report Schedule B”, “PM Report Schedule C” and “PM Report Schedule D”
|
| 18 |
June 10, 2014 |
In addition to general procedure updates, the following major changes were made: |
| Readme : Updated “Language Policy” |
| Safety : Updated “Important...X-Ray Protection” |
| Part List : Added “Gantry Parts List - Cj M40 RoHS”, “Console Parts List - Cj M40 RoHS”, “NGPDU Part List - Cj M40 RoHS”, “Table Parts List - Cj M40 RoHS”, “Misc. Items Parts List - Cj M40 RoHS” |
Replacement :- Gantry > DAS (SDCB) : Updated “SDCB Backplane Replacement”
- Gantry > DAS (Merc40) : Updated “Detector Module Replacement - Merc40”
- Table > GT1700V Table : Updated “Cradle Absolute Encoder Replacement”, “Cradle Replacement”
|
| 17 |
April 23, 2014 |
In addition to general procedure updates, the following major changes were made: |
| Safety : Updated “ESD Grounding Locations” |
| Software > Software Installation Procedures (LFC) : Updated “(13HW38.14) Service Pack Installation Procedure” |
| Troubleshooting > Image Quality : Updated “Scan Analysis (Merc40)” |
| 16 |
April 02, 2014 |
In addition to general procedure updates, the following major changes were made: |
| Software > Software Installation Procedures (LFC) : Updated “(13HW38.14) Service Pack Installation Procedure” |
| 15 |
March 20, 2014 |
In addition to general procedure updates, the following major changes were made: |
| Readme : Updated “CT/PET Common Glossary” |
| Safety : Updated “Equipment Service - Gantry” |
| Installation : Updated “Optima CT660 Installation” and “Installation Forms” |
Functional Checks :- Added “Enclosure Leakage (Patient Touch) and System Chassis Leakage Tests”
- > Gantry : Updated “zalignmentWizard User Instructions (Merc40)”
|
| Software > Software Installation Procedures (LFC) : Updated “(11HW34.x / 12HW12.x) RIO LFC Quick Guide”, “(13HW38.x) LFC Quick Guide”. Added “(12HW28.8) Service Pack 3 Installation Procedure” |
Theory :- > Gantry : Added “CDAS Theory” and “Gantry Thermal Theory (Merc40)”
- > Gantry > DAS and Detector : Added “CDAS Theory” and “Merc40 Detector Theory”
- > System : Updated “Energy Saving Mode”
|
Troubleshooting :- > Gantry : Added “SDCB DAS Troubleshooting”, “Gantry Thermal Control Troubleshooting (Cj M40)”
- > Image Quality : Added “Scan Analysis (Merc40)”
|
Replacement :- Updated “Certified Component Verification Form e4879 - Cj”
- > Console (NIO64) : Added “GSCB Replacement (NIO64)”
- > Gantry : Added “Simplified HVDC Power Pan ASM Replacement”
- > Gantry > DAS and Detector Common : Updated “FRDM Replacement Procedure”
- > Gantry > DAS (Merc40) : Updated “Detector Air Plenum Removal and Installation”, “Detector Carbon Window Replacement - Merc40”, “Detector Collimator Unit Replacement - Merc40 “, “Detector Module Replacement - Merc40”, “Thermistor Assembly Replacement”
- > Gantry > Enclosure : Updated “Gantry Heater Replacement”. Added “Gantry Top Fan Replacement”
- > Gantry > ORP and TGPU : Added “ORP Replacement (M40)”
|
| 14 |
January 29, 2014 |
In addition to general procedure updates, the following major changes were made: |
| Safety : Updated “Equipment Service - Console” |
Installation :- Updated “Optima CT660 Installation” and “Installation Forms”
- > Option : Updated “Accessory Interface Hardware Installation (Dir 5445931-1EN)”, “AWE Connection Option Installation”, “Xtream Injector Option”. Added “NIO Console and GOC6.6 Console Long Cable Kit Installation Manual”
|
| Align, Setup, Calibrations > Console : Added “Z820 Host PC BIOS Setup “, “GSCB Configuration and Intercom Setup”. Updated “Z800 Host Computer BIOS setting” |
| Functional Checks > Console : Added “Peripheral Tower Functional Test (PMT-20)” |
| Software > Software Installation Procedures (LFC) : Updated “Install Software Option”, “(13HW38.x) Load From Cold - Optima CT660”. Added “(13HW38.14) Service Pack Installation Procedure”. |
| System Diagrams : Updated “Optima CT660 Functional Interconnects ” |
Troubleshooting :- > Console : Updated “Misc. Troubleshooting Information”
- > Diagnostics : Updated “DAS Tools (Merc40)”
|
| Part List : Updated “Gantry Parts List”, “Console Parts List”, “Misc. Items Parts List”. |
Replacement :- > Gantry > DAS (Merc40) : Updated “CDCB Replacement - Merc40”, “CFHC Replacement”, “Detector Air Plenum Removal and Installation”, “Detector Module Replacement - Merc40”, “Power Board Replacement - Merc40”
- > Gantry > High Voltage : Updated “Auxiliary Assembly Replacement”, “HEMIT Tank Replacement”, “Power Unit Replacement”.
|
| 13 |
January 07, 2014 |
In addition to general procedure updates, the following major changes were made: |
| Readme : Updated “Revision History” and “CT/PET Common Glossary” |
| Safety : Updated ”Equipment Service - Gantry“, “Equipment Service - Lockout-Tagout-PPE”, “Equipment Service - NGPDU”, “Important...X-Ray Protection” and “Normal System Operational Safety” |
| Installation : Updated “Optima CT660 Pre-Installation” and “Optima CT660 Installation” |
Align, Setup, Calibrations :- > Console : Updated “Data Privacy (EA3 User Authorization) Configuration” and “Dose Check Management and EA3 Configuration”. Added “System Configuration - NIO64”
- > Gantry > Enclosure : Updated “Alignment Light Adjustment” and “Alignment Light Characterization”
|
| Functional Checks > Gantry : Updated “Gantry Rotation Safety Check”. Added “zalignmentWizard User Instructions - Merc40” |
| Software > Software Installation Procedures (LFC) : Added “(13HW38.x) Load From Cold - Optima CT660” and “(13HW38.x) LFC Quick Guide” |
| Theory > Console : Added “NIO64 Console Theory” and “LCD Video Monitor Setup” |
| System Diagrams : Updated “Optima CT660 Functional Interconnects for Cj M40” |
Troubleshooting :- > Console : Added ”AC Box Troubleshooting (5412524-2)”, “GOC6.6 Console Network Switch Troubleshooting”, “GOC6.6 / NIO64 Misc. Troubleshooting Information” and “Z820 Host PC Troubleshooting”. Updated “Real-In-On Console Diagnostics (10HW10.x / 10HW33.x / 11HW34.x)”.
- > Diagnostics : Added ”DAS Tools (Merc40)” and “Scan Data Path Diagnostic (Merc40)”. Updated “Scan Data Path Diagnostic”
- > Image Quality : Updated “Scan Analysis”
- > Table : Updated “GTCB Error Code List (in gesyslog)”
- > X-Ray Generation : Added “Rotor Control tool with LB tube”
|
| Part List : Updated “Gantry Parts List” |
Replacement :- Updated “Fuse Replacement Procedures”
- > Console : Updated “HP Host Computer Z800 FX1800 Graphic Card Replacement”
- > Gantry > Collimator : Updated “Collimator Filter Removal and Replacement”
- > Gantry > DAS and Detector Common : Updated “FRDM Replacement Procedure”
- > Gantry : Added DAS (Merc40)
- > Gantry > Enclosure : Updated “Gantry Front Cover Removal and Re-Install” and “Gantry Rear Cover Removal and Re-Install”
- > Gantry > Tilt : Updated “Tilt Potentiometer Replacement”
- > Gantry > Tube : Updated “X-Ray Tube Replacement”
|
| 12 |
July 24, 2013 |
In addition to general procedure updates, the following major changes were made: |
| Readme : Updated “Revision History” |
Align, Setup, Cals :- > Console : Updated “LCD Video Monitor Setup”
- > Gantry > Tube : Updated “Cold ISO Alignment” and “Z-Alignment”
|
| Software > Software Installation Procedures (LFC) : Updated “(12HW28.x) Load From Cold - Optima CT660”, “(12HW28.x) RIO LFC Quick Guide” and “(12HW28.8) Service Pack Installation Procedure” |
| Part List : Updated “Gantry Parts List” |
Replacement :- > Console : Updated “HP Host Computer Z800 Scanfile HDD Replacement”
- > Gantry > Display : Added “Xtream Display CF Memory Card Replacement”
|
| Planned Maintenance and Tasks : Updated “Planned Maintenance Schedule A”, “Planned Maintenance Schedule B”, “Planned Maintenance Schedule C” and “Planned Maintenance Schedule D” |
| 11 |
February 25, 2013 |
In addition to general procedure updates, the following major changes were made: |
| Software > Software Installation Procedures (LFC) : Added “(12HW12.5) Service Pack Installation Procedure” |
| 10 |
February 22, 2013 |
In addition to general procedure updates, the following major changes were made: |
| Readme : Updated “Revision History” |
| Safety : Updated “Equipment Service - NGPDU” and “Torque Wrench Information” |
| Align, Setup, Cals > Table > Lite Table : Updated “Parallel Setting of Cradle” and “Short Footprint Setting Procedure (BSD Select)” |
| Software > Software Installation Procedures (LFC) : Added “(12HW28.8) Service Pack Installation Procedure” |
Replacement :- > Table > Lite Table : Updated “Hydraulic Cylinder Replacement” and “Pump Assembly Replacement”
- > Gantry > DAS (SDCB) : Added “SDCB Fiber Optic Transceiver Replacement”
|
| 9 |
August 28, 2012 |
In addition to general procedure updates, the following major changes were made: |
| Readme : Updated “Revision History” and “Damage in Transportation” |
| Safety : Updated “Equipment Service - FRU Parts Management” |
Installation :- Updated “Pre-Installation” and “Installation”
- > Option : Added “Accessory Interface Hardware Installation (Dir 5445930-1EN)”, “ASiR Option Manual (Dir 5436607-1EN)”, “ASiR Upgrade Manual (Dir 5451977-1EN)” and “Phase 2.5 Product Grade Collector Installation (Dir 5456075-1EN)”. Updated “AWE Connection Option Installation” and “Remote Monitor Installation (Dir 2390072-100)”
|
Align, Setup, Cals :- > Gantry : Updated “Intercom Adjustment” and “Slip Ring Adjustments”
- > Gantry > Tube : Updated “Cold ISO Alignment” and “Z-Alignment”
|
| Software > Software Installation Procedures (LFC) : Added “(12HW28.x) Load From Cold - Optima CT660“ and “(12HW28.x) RIO LFC Quick Guide” |
| Theory > Gantry > DAS and Detector : Updated “VCT Detector Theory” and “VDAS and HDAS Theory” |
| Troubleshooting > Image Quality : Updated “Scan Analysis (SDCB)” |
| Part List : Updated “Console Parts List” |
Replacement > Gantry :- > Collimator : Updated “Cam Drive Motor and Flex Coupling Replacement” and “Collimator Assembly Replacement”
- > Enclosure : Updated “Intercom Board Replacement”
- > Slipring : Updated “Slip Ring Platter Replacement”
- >Tube : Updated “X-Ray Tube Replacement”
|
| 8 |
April 17, 2012 |
Readme : Updated “Revision History” and “Damage in Transportation” |
Installation :- Updated “Pre-Installation” and “Installation”
- > Option : Updated “AWE Connection Option Installation”
|
Align, Setup, Cals :- > Console : Updated “Evaluation of Dose SR Compatibility” and “Host Computer BIOS setting”
- > Gantry > Enclosure : Added “Auto Alignment Light Setting”
|
| Functional Checks : Updated “Quality Assurance Test” |
| Software > Software Installation Procedures (LFC) :Updated “(11HW34.x / 12HW12.x) Load From Cold - Optima CT660“, “(11HW34.x / 12HW12.x) RIO LFC Quick Guide “ and “Service Pack Installation Procedure” |
| Theory > Table : Updated “Quality Assurance Test” |
Replacement :- > Gantry : Updated “IPC Board Replacement”
- > Gantry > Collimator : Updated “Cam Drive Motor and Flex Coupling Replacement” and “Collimator Assembly Replacement”
- > Gantry > DAS and Detector Common : Updated “FRDM Replacement Procedure”
- > Table > GT2000 Table : Updated “Cradle Lock Assy Replacement”, “Cadle Motor and Clutch Asy Replacement”, “Cradle Replacement”, “GTCB Replacement”, “CTCN Replacement”, “Linear Scale Assy Replacement”, “Rear Touch Sensor Plate Replacement”, “Touch Sensor L Assy Replacement” and “Touch Sensor R Assy Replacement”
|
| 7 |
February 07, 2012 |
Readme : Updated “Revision History” |
| Installation > Option : Updated “AWE Connection Option Installation” |
| Functional Checks > Gantry : Updated “HV Tank Feedback Resistor Verification” and “HHS Scans” |
| Software > Software Installation Procedures (LFC) :Updated “(10HW33.x / 10HW02.x) RIO LFC Quick Guide“, (10HW33.x) Service Pack Installation Procedure” and “(11HW34.x) Load From Cold - Optima CT660”. Added “(11HW34.x) RIO LFC Quick Guide” and “(11HW34.x) Service Pack Installation Procedure” |
| Theory > Gantry : Updated “HDAS with SDCB Theory” |
| System Diagrams : Updated “Optima CT660 Functional Interconnects” |
| Part List : Updated “Gantry Parts List” |
Replacement > Gantry :- > DAS (SDCB) : Updated “SDCB Replacement” and “SDCB Backplane Replacement”
- > High Voltage : Updated “Auxiliary Assembly Replacement”, “HEMIT Tank Replacement”, “HV Tank Replacement”, “Power Unit Replacement” and “Programmed KV Control Board Replacement”
- > Tube : Updated “X-Ray Tube Replacement”
|
Planned Maintenance and Tasks :- Updated “Planned Maintenance Schedule A”, “Planned Maintenance Schedule B”, “Planned Maintenance Schedule C” and “Planned Maintenance Schedule D”
- > PM Data Reports : Updated “PM Report Schedule D”
|
| 6 |
November 22, 2011 |
Readme : Updated “Revision History” |
Installation :- Updated “Pre-installation” and “Installation”
- > Option : Added “2 Phase UPS Installation Supplement (Dir 5428231-1EN)”, “AWE Connection Option Installation” and “Gantry Rear Control Panel Installation Manual (Dir 5426891-1EN)”
|
Align, Setup, Cals :- > Console : Added “Host Computer BIOS setting”. Updated “System Configuration - RIO”
- > Gantry > Enclosure : Added “Xtream Display Default Patient Positioning” and “Xtream Display Patient Information”
|
Software :- > Software Installation Procedures (LFC) : Added “11HW Load From Cold - Optima CT660”. Updated “System State Save Restore”, “10HW02.x / 10HW33.x Load From Cold - Optima CT660”, “(11HWxx.xx / 10HW33.x / 10HW02.x) RIO LFC Quick Guide“, “Install Software Options - Optima CT660“ and “Service Pack Installation Procedure”
- > System Utilities : Updated “Linux Commands”
|
Theory :- > Console : Updated “Real-In-One Console Theory” and “DVD Tested Media “
- > Gantry : Updated “Xtream Display Theory “
- > Gantry > DAS and Detector : Added “HDAS with SDCB Theory”
|
| System Diagrams : Updated “Optima CT660 Functional Interconnects” |
Troubleshooting :- > Console : Updated “Real-In-On Console Diagnostics”
- > Diagnostics : Added “DAS Tools (HDAS with SDCB)“ and “Scan Data Path Diagnostic (HDAS with SDCB)”
- > Image Quality : Updated “Quick Snap and IQ Snap Data Archive and Shipping Procedures “ and “Scan Analysis”
|
Replacement :- > Gantry : Added "DAS (HALO) Contents” and “DAS (SDCB) Contents”
- > Gantry > Enclosure : Added “Gantry Microphone Replacement “ and “Usage of the Support Tool for Gantry Bore Cover “
- > Gantry > Slipring : Updated “Slipring Brush Block Replacement “
- > Table > GT2000 Table : Updated “Cradle Encoder Spool Wire Replacement “
- > Table > GT1700V Table : Updated “Cradle Encoder Spool Wire Replacement “
|
Planned Maintenance and Tasks :- Updated ”Planned Maintenance Schedule C”
- > PM Data Reports : Updated “PM Report Schedule C”
|
| 5 |
August 04, 2011 |
Readme : Updated “Revision History” and “CT/PET Common Glossary” |
| Align, Setup, Cals > Gantry > Tilt : Updated “Hydraulic Tilt Motor Fluid Check and Fill Procedure” |
| Functional Checks : Updated “System Scanning Test” |
Software > Software Installation Procedures (LFC) :- Updated “10HW02.x / 10HW33.x Load From Cold - Optima CT660”
- Added “Service Pack Installation Procedure”
|
| Troubleshooting > Image Quality : Updated “Streak Test” |
| Replacement > Gantry > Das and Detector : Updated “Detector Thermistor Replacement” and “FRDM Replacement Procedure” |
| 4 |
February 28, 2011 |
Readme : Updated “Revision History” |
Installation :- Updated “Pre-installation” and “Installation”
- > Option : Added “Option Bare media handling”. Updated “DVD RAM Cartridge Removal”.
|
Align, Setup, Cals :- > Console : Updated “System Configuration (reconfig) - RIO”
- > Table : Added “GT2000 Table Contents”
|
| Software > Software Installation Procedures (LFC) : Updated “System State Save Restore “, “(10HW02.x / 10HW33.x) RIO LFC Quick Guide” and “(10HW02.x / 10HW33.x) RIO LFC Quick Guide” |
| Theory > Table : Added “GT2000 Table Theory” |
| System Diagrams : Updated “Optima CT660 Functional Interconnects” |
| Troubleshooting > Console : Added “RAID Troubleshooting” |
| Part List : Updated ”Console Parts List”, “Gantry Parts List” and “Table Parts List” |
Replacement :- > Console : Updated “HP Host Computer Z800 RAID Controller”
- > Table : Added “GT2000 Table Contents”
|
| 3 |
September 06, 2010 |
Readme : Updated “Revision History” |
Installation :- Updated “Optima CT660 Installation”
- >Option : Added “Power Option Installation”
|
| Align, Setup, Cals > Cosole : Updated “System Configuration - RIO” |
Functional Checks :- >Console : Updated “Peripheral Tower Functional Test”
- >Gantry : Updated “HHS Scan”
|
| Software > Software Installation Procedures (LFC) : Updated “System State Save Restore”, “10HW02.x Load From Cold - Optima CT660”, “(10HW02.X) RIO LFC Quick Guide” and “Install Software Options - Optima CT660” |
| Part List : Updated “Console Parts List” |
| Planned Maintenance and Tasks : Updated “Planned Maintenance Schedule D” |
| 2 |
April 19, 2010 |
Safety : Updated “Equipment Service - Console” and “Equipment Service - Lockout-Tagout-PPE” |
| Installation : Updated “Optima CT660 Installation” |
Align, Setup, Cals :- >Gantry : Updated “Gantry Balance Procedure”
- >Gantry > Tube : Updated “Cold ISO Alignment” and “Tube Install Certification”
|
| Functional Checks > Gantry : Updated “Gantry and Detector Filter Cleaning” |
| Software > Software Installation Procedures (LFC) : Updated “10HW02.x Load From Cold - Optima CT660” and “System State Save Restore” |
| Theory > Gantry > DAS and Detector : Updated “VDAS and HDAS Theory” |
| System Diagrams : Updated “Optima CT660 Functional Interconnects” |
| Part List : Updated “Console Parts List” |
Replacement :- Added “Certified Component Replacement Retest Matrix” and “Certified Component Verification Form e4879”
- >Console : Updated “DVD Peripheral Tower Replacement” and “LCD Monitor Replacement”
- >Gantry : Updated “AC Power Distribution and CB Assembly Replacement”
- >Gantry > Display : Updated “Breathing Light Assembly Replacement”
- >Gantry > Enclosure : Updated “Gantry Bore Cover Removal and Re-Install” and “Gantry Front Cover Removal and Re-Install”
- >Gantry > High Voltage : Updated “HEMIT Tank Replacement”
- >Gantry > ORP and TGPU : Updated “ORP Replacement”
- >Gantry > Tube : Updated “X-Ray Tube Replacement”
|
| Planned Maintenance and Tasks : Updated “Planned Maintenance Schedule A, B, C and D” |
| 1 |
March 24, 2010 |
Initial Release |