• Topic ID: concept_pxk_rgh_5rb
  • Version: 2.0
  • Date: Feb 22, 2022 12:44:43 AM

(15HW25.2) Service Pack 3.3 Installation Procedure

1 Personnel Requirements

2 Procedure Overview

This procedure describes 15HW25.2 Service Pack 3.3 installation procedure.

The Service Pack 3.1 software CDROM contains Service Pack 3.0, 3.1 and Service Pack 3.3 patches.

3 Preliminary Requirements

3.1 Tools and Test Equipment

No tools or test equipment required

3.2 Consumables

No consumables required

3.3 Replacement Parts

No replacement parts required

3.4 Safety

No general safety infomation

3.5 Required Conditions

4 Procedure

4.1 Check software version currently installed

Confirm that the base software version is 15HW25.2 with Service Pack3.1 as follows.

  1. Open a Unix shell.

  2. Type swhwinfoENTER

  3. Confirm that the software version is 15HW25.2_SP3-0-1 <hw info (ex. H40_P2_SS64_G_GTL)>.

    If not, upgrade the system at first.

4.2 Service Pack Installation

  1. If application software is up, perform Application Shutdown from Common Service Desktop.

  2. Open a Unix shell.

  3. In the Unix Shell, become root:

    1. {ctuser@hostname} su -

    2. Password: <password>

  4. Insert the Service Pack CD into DVD drive in the host.

  5. Perform the following steps to install the Service Pack:

    note:

    Always install Service Pack 3.0. 3.1 and Service Pack 3.3 together.

    1. [root@hostname]# patch_install -c

    2. The following message is displayed.

      note:

      The message may be different from below depending on the service pack already installed.

      These patches from cdrom will be installed:

      15HW25.2_GEHCServicePack3.0

      15HW25.2_GEHCServicePack3.1

      15HW25.2_GEHCServicePack3.3

      -------------------------------------------

      I will install update 15HW25.2_GEHCServicePack3.0 from CDROM, is this ok ? [y/n]

    3. Input y to install Service Pack 3.0.

    4. Confirm “Update 15HW25.2_GEHCServicePack3.0 has been installed successfully” message.

    5. The following message is displayed.

      note:

      The message may be different from below depending on the service pack already installed.

      Update 15HW25.2_GEHCServicePack3.0 has been installed successfully.

      -------------------------------------------

      I will install update 15HW25.2_GEHCServicePack3.1 from CDROM, is this ok ? [y/n]

    6. Input y to install Service Pack 3.1.

    7. Confirm “Update 15HW25.2_GEHCServicePack3.1 has been installed successfully” message.

    8. The following message is displayed.

      I will install update 15HW25.2_GEHCServicePack3.3 from CDROM, is this ok ? [y/n]

      Input y to install Service Pack 3.3.

    9. Confirm “Update 15HW25.2_GEHCServicePack3.3 has been installed successfully” message.
  6. Check Software Version:

    1. Type: [root@hostname]# swhwinfo ENTER

    2. System should reply with a response similar to the following:

      15HW25.2_SP3-3-2 <hw info (ex. H40_P2_SS64_G_GTL) >

  7. Remove the Service Pack CD from Host DVD drive.

  8. Reboot the system:

    [root@hostname]# reboot

  9. Perform Flash Download.

5 Finalization

  1. Perform sanity scans (scout, Helical, Axial, Cine).

  2. Confirm that all scans are successful and images are displayed.

6 Appendix - How to uninstall Service Pack

  1. If the following software options are installed, uninstall them at first:

    • DAP -DigtalProcessing

    • DAP -ASiR-V

    • DAP -MAR

  2. If Application software is up, perform Application Shutdown from Common Service Desktop.

  3. Open a Unix shell.

  4. In the Unix Shell, become root:

    1. {ctuser@hostname} su -

    2. Password: <password>

  5. Perform the following steps to uninstall the Service Pack CD:

    1. [root@hostname]# patch_install -u

    2. The following message is displayed.

      The following patches are installed on system:

      ----------------------------------------------

      15HW25.2_GEHCServicePack3.3

      15HW25.2_GEHCServicePack3.1

      15HW25.2_GEHCServicePack3.0

      15HW25.2_GEHCServicePack2.0

      15HW25.2_GEHCServicePack1.0

      ----------------------------------------------

      I will uninstall update 15HW25.2_GEHCServicePack3.3, is this ok? [y/n]

      Input y to uninstall Service Pack 3.3.

      I will uninstall update 15HW25.2_GEHCServicePack3.1, is this ok? [y/n]

    3. Input y to uninstall Service Pack 3.1.

    4. Confirm the following message.

      Update 15HW25.2_GEHCServicePack3.1 has been uninstalled successfully.

    5. The following message is displayed.

      ###############################################################

      It's necessary to uninstall SP3.0 after uninstalling SP3.1.

      ###############################################################

      There are some running programs that use files deleted by recent upgrade.

      You may wish to restart some of them. Run 'zypper ps' to list these programs.

      Update 15HW25.2_GEHCServicePack3.1 has been uninstalled successfully.

      ----------------------------------------------

      I will uninstall update 15HW25.2_GEHCServicePack3.0, is this ok? [y/n]

    6. Input y to uninstall Service Pack 3.0.

    7. Confirm the following message.

      Update 15HW25.2_GEHCServicePack3.0 has been uninstalled successfully.

  6. The following message is displayed. Input n to quit patch uninstall menu.

    Update 15HW25.2_GEHCServicePack3.0 has been uninstalled successfully.

    ----------------------------------------------

    I will uninstall update 15HW25.2_GEHCServicePack2.0, is this ok? [y/n]

    1. Input n to uninstall Service Pack 2.0.

    2. Confirm the following message.

      Exiting on user command

      I will uninstall update 15HW25.2_GEHCServicePack1.0, is this ok? [y/n]

    3. Input n to uninstall Service Pack 1.0

  7. Reboot the system:

    [root@hostname]# reboot

  8. Perform Flash Download.

  9. Perform sanity scans. Refer to Finalization for the details.