(15HW25.2) Service Pack 2 Installation Procedure
1 Personnel Requirements
2 Overview
This procedure describes 15W25.2 Service Pack 2 installation procedure.
The Service Pack 2 software CD contains Service Pack 1 and 2 patches.
3 Preliminary Requirements
3.1 Tools and Test Equipment
No tools or test equipment required
3.2 Consumables
No consumables required
3.3 Replacement Parts
No replacement parts required
3.4 Safety
No general safety infomation
3.5 Required Conditions
4 Procedure
4.1 Check software version currently installed
-
Open a Unix shell.
-
Type swhwinfoENTER
Confirm that the base software version is 15HW25.2.
4.2 Service Pack Installation
-
Perform Save System State (select [All] button).
-
If Exam Split option is installed, record the mode (HES or VES) by the following command.
-
Open Unix shell and and log on as root.
-
Type ls -l ~ctuser/ves/.hesMode
-
If there is .hesMode file, it is HES mode. If not, It is VES mode.
-
-
Uninstall the following options if they are installed. And record them.
-
AutoBone_Xpress
-
AVA_Xpress
-
CardEP
-
CardIQ_Xpress_Reveal
-
CT_Colono_Pro3D_EC
-
CT_Perfusion_4D_MultiOrgan
-
CT_Perfusion_4D_Neuro
-
Volume Viewer
-
-
Perform Application Shutdown from the CSD Utilities menu.
-
Insert the Service Pack CD into the Host DVD drive
-
Open a Unix shell.
-
In the Unix Shell, become root:
-
{ctuser@hostname} su -
-
Password: <password>
-
-
Perform the following steps to install the Service Pack CD:
-
[root@hostname]# patch_install -c
note:If Service Pack 1.0 is already installed, go to step e.
-
The following message is displayed.
The following files found in cdrom match current system:
==============================================================
15HW25.2_GEHCServicePack1.0-1-1.i386.rpm
15HW25.2_GEHCServicePack2.0-1-1.i386.rpm
==============================================================
copy patches to system...
These patches from cdrom will be installed:
15HW25.2_GEHCServicePack1.0
--------------------------------------------------------------
I will install update 15HW25.2_GEHCServicePack1.0 from CDROM, is this ok ? [y/n]
-
Input y to install Service Pack 1.0.
-
Confirm that the following message is displayed:
Update 15HW25.2_GEHCServicePack1.0 has been installed successfully.
-
The following message is displayed.
I will install update 15HW25.2_GEHCServicePack2.0 from CDROM, is this ok ? [y/n]
-
Input y to install Service Pack 2.0.
-
Confirm that the following message is displayed:
Update 15HW25.2_GEHCServicePack2.0 has been installed successfully.
-
-
Check Software Version:
[root@hostname]# swhwinfo ENTER
-
System should reply with a response similar to the following:
15HW25.2_SP2-0-1. <hw info (ex. H40_P2_SS64_G_GTL)>
-
Remove the Service Pack CD from Host DVD drive
-
Reboot the system:
[root@hostname]# reboot
-
Perform Restore System State. (Select [All] button)
-
Restart the system:
-
Confirm all removed option keys are restored correctly by Verify Options in Common Service Desktop.
5 Finalization
No finalization steps.
6 Appendix - How to uninstall Service Pack
-
Perform Save System State (select [All] button)
-
If Exam Split option is installed, record the mode (HES or VES) by the following command.
-
Open Unix shell and and log on as root.
-
Type ls -l ~ctuser/ves/.hesMode
-
If there is .hesMode file, it is HES mode. If not, It is VES mode.
-
-
Uninstall the following options if they are installed.
-
AutoBone_Xpress
-
AVA_Xpress
-
CardEP
-
CardIQ_Xpress_Reveal
-
CT_Colono_Pro3D_EC
-
CT_Perfusion_4D_MultiOrgan
-
CT_Perfusion_4D_Neuro
-
Volume Viewer
-
-
Perform Application Shutdown from the CSD Utilities menu.
-
Open a Unix shell.
-
In the Unix Shell, become root:
-
{ctuser@hostname} su -
-
Password: <password>
-
-
Perform the following steps to install the Service Pack CD:
-
[root@hostname]# patch_install -u
-
The following message is displayed.
The following patches are installed on system:
--------------------------------------------------------------
15HW25.2_GEHCServicePack2.0
15HW25.2_GEHCServicePack1.0
--------------------------------------------------------------
I will uninstall update 15HW25.2_GEHCServicePack2.0, is this ok? [y/n]
-
Input y to uninstall Service Pack 2.0.
-
The following message is displayed.
The following package is going to be REMOVED:
15HW25.2_GEHCServicePack2.0
...
Update 15HW25.2_GEHCServicePack2.0 has been uninstalled successfully.
I will uninstall update 15HW25.2_GEHCServicePack1.0, is this ok? [y/n]
-
Input y to uninstall Service Pack 1.0.
-
The following message is displayed.
The following package is going to be REMOVED:
15HW25.2_GEHCServicePack1.0
...
Update 15HW25.2_GEHCServicePack1.0 has been uninstalled successfully.
-
-
Check Software Version:
[root@hostname]# swhwinfo ENTER
-
System should reply with a response similar to the following:
15HW25.2. <hw info (ex. H40_P2_SS64_G_GTL)>
-
Reboot the system:
[root@hostname]# reboot
-
Restore System State.
-
Restart the system.
-
Confirm all removed option keys are restored correctly by Verify Options in Common Service Desktop.