• Topic ID: id_16157515
  • Version: 4.0
  • Date: Jan 20, 2020 8:34:56 PM

Tilt Interference and Limit Switches Replacement

Prerequisites

Overview

Procedure

  1. Remove gantry covers as required.

    Refer to Parts Replacement > Gantry > Enclosure > (Cover Removal Procedure).

  2. Stop the rotor of X-ray tube in case of Liquid Bearing Tube before HVDC off. Refer to Liquid Bearing Tube Rotor stop procedure for details.
  3. Disconnect molex connector to replace harness.
  4. Cut ty-raps as necessary.
  5. Reassemble gantry.

Finalization

  1. Verify full range of tilt operation.