- Topic ID: id_11038811
- Version: 4.0
- Date: Nov 27, 2020 2:13:28 AM
Slipring Receiver (850 Mbaud) Replacement
Prerequisites
This procedure describes the Slipring Receiver replacement.
Procedure
- Move table to its lowest elevation.
- Remove rear gantry cover Gantry Rear Cover Removal and Re-install.
- Remove J1 and J2 cables.
- Remove hex screws holding the receiver to the antenna.
Figure 1. 850 MBaud Slipring Receiver

- Replace receiver.
- Verify antenna gap and alignment adjustments.
Finalization
- Hardware Reset
- Acquire 10 scouts: (120kV/40mA, 1000mm table movement)
- Acquire 100 axials: (120kV/80mA, 0.5 sec. Scan)
- Acquire 1 helical: (120kV/40mA, 30 sec. Scan),
- Acquire 10 axial scans: (120kV/400mA, 4 sec. Scan)
- Verify NO increase in corrected or uncorrected FEC errors.
- Reset DIP Stats: Perform the following procedure to reset cumulative
DIP Stats (shown on the Common Service Desktop Home Page):
- Open a Unix shell.
- Type: cd /usr/g/service/log
- Type: mv dip.stats dip.stats.old
- Shutdown and restart the system. This will reinitialize the dip.stats file.
- Save System State.