- Topic ID: id_16157871
- Version: 4.0
- Date: Jan 20, 2020 8:34:15 PM
Linear Scale Tape Replacement
Prerequisites
Overview
This procedure is for the Global PET/CT Table, with fixed position IMS and GT1700V Table.
Procedure
- Raise the Table to maximum height.
- Move the Cradle to OUT limit position.
- Remove power from Table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
- Remove the following Table component and covers:
-
Cradle
-
Top Cover (Right/Left)
-
Rear Top Cover
-
Rail Cover (Left)
Figure 1. Table Covers

-
- Remove fixing bracket from the Table frame (see Figure 2).
Figure 2. Fixing Bracket Removal

- Peel off the linear scale tape from the groove on the inside of the Table frame, and take off the adhesive tape in that place.
- Clean and dirt off the attaching surface of the groove using
an alcohol.
Figure 3. Linear Scale Tape Removal

- notice
- Perform this step depending on the following Table type:
(For Global PET/CT Table, with fixed position IMS) Skip this step.
(For GT1700V Table) Cut new linear scale tape into 1895 ± 5 mm length.
- Peel off a paper from the new linear scale tape, and attach
it along the under edge of the groove (see Figure 3).note:
If the front cradle stopper becomes a hindrance to attaching the linear scale tape, remove it.
![]() |
Finalization
- Re-install the cradle.
- Power up the Table from the Service Switch Panel.
- Verify that the cradle IN/OUT function is operating normally.
- Turn off all 3 switches (Axial Drive, HVDC, 120VAC), and re-install the Table covers.