HP Host Computer Z800 KDIP Card Replacement

Prerequisites

Procedure

  1. Select one of the following methods to Power OFF the Operator Console:
    1. If Applications are up, click on the Shut Down button and select Shutdown.

      The Operator Console monitor will display a 'Power Down' message when it is acceptable to power OFF the Operator Console.

    2. If Applications are down, open a Unix Shell. Type: halt.
  2. Power OFF the console at the front panel switch.
  3. Remove the left side cover from the console.

    Refer to Replacement > Console > (Console Cover Removal and Installation Procedure).

  4. Remove the following components from the host computer (Z800):
    • Side Access Panel

    • Expansion Card Support

  5. Replace the existing z800 KDIP card with new one.
  6. Reinstall all removed components, and reconnect any cables that have been disconnected.

Finalization

  1. Refer to System Scanning Test to confirm proper operation.
  2. Reinstall the console covers.