- Topic ID: id_11038867
- Version: 3.0
- Date: Apr 9, 2020 8:42:55 PM
HHS Scans
Prerequisites
The objective of this procedure is to validate HHS scans fit within specifications:
-
Attach oscilloscope to Inverter
-
Set up oscilloscope
-
Take each scan and record measurements
Procedure
- Remove gantry right side cover.
- Turn OFF all 3 switches (AXIAL DRIVE ENABLE, HVDC ENABLE, 120 VAC) on the Service Switch Panel.
- Remove and set aside the top and front gantry covers.
- Rotate the gantry such that HP inverter assembly is in the 2 o'clock position.
- Lock the gantry in position, using the rotational lock. Ensure that gantry rotation is locked by attempting to rotate the gantry by hand.
- Remove the large Inverter covers.
- Attach a protective ESD wrist strap and use while making connections to the HP inverter test points in the next step.
- Setup the oscilloscope for your first scan
- To minimize noise and obtain a common ground, you should plug the oscilloscope into either the gantry or the console. If you do not, this test won't work.
- Connect channel one probe to KV_N (TP6) and ground to KV_GND (TP8).
- Connect channel two probe to MA_N (TP5) and ground to MA_GND (TP7).
- Set horizontal scale = 20 ms, vertical scale = 1V/div
- Trigger channel one, negative slope, DC couple, trigger mode normal.
- Launch the KV MA TOOL.
- Begin scans sequence. For all scans gantry rotation should be disabled. See the HHS data sheet for the list of scans you will need to take.
- Using cursors, record the mean value for mA between 4 and 6 seconds. Multiply this value by Record this value in the mA @ 5mS column on the HHS scans data sheet.
- Using cursors, record the mean value for kV between 10-20 seconds. Record this value in the kV column on the HHS data sheet.
- Record the time and the mA value listed in the corresponding column on the HHS scans data sheet for each scan.
- Record the actual kV value at the bleeder.
- Continue for all scans. You will need to adjust the vertical scale accordingly on the scope as the scan kV increases.
- Record all results on the HHS Scans Data Sheet. You will need to print these sheets.
Finalization
- Ensure all measurements are within the specified limits. If any one scan is out of specification limits, take the appropriate service action.