Gantry Rotation Safety Check
Prerequisites
Overview
Use DDC to Rotate the Gantry for 10 minutes, at the fastest allowed gantry speed.
Procedure
- Make sure the Axial Drive Enable switch (S2) and HVDC swich (S3) on the service switch panel are off.
- Manually rotate the gantry through a full rotation listening
for an contact between stationary and rotating parts
-
Listen for an contact between stationary and rotating parts. Correct any issues found before continuing.
-
Watch and listen for any loose parts/cables and tighten them before continuing.
-
- Turn on the HVDC and Axial Drive service switches.
- Start the Axial Functional diagnostic.
- Open the Common Service Desktop
- Select the Diagnostics Icon
- Select the Axial Functional menu item to start the diagnostic.
- notice
- Select a rotation speed of 2 seconds and run the test. This
step should be run for a minimum of 30 seconds.
Listen for an contact between stationary and rotating parts. Correct any issues found before continuing.
note:Be ready to click on the Stop button on the menu immediately if you hear any problems. (This stops the gantry faster than the red emergency stop (e-stop) button)
- Repeat Step 5 for rotation speeds of 1.0 and 0.5 seconds.note:
(For BrightSpeed Select) rotation speeds of 1.0 and 0.8 seconds.
- Wait 3 minutes after the gantry stops rotating before continuing
to allow for axial drive cooling.
Be ready to click on the Stop button on the menu immediately if you hear any problems. (This stops the gantry faster than the e-off button)
- If no issues arise, gantry rotation capability is safe for use.
![]() |
Finalization
No finalization steps.