GDAS 48V Power Supply Checks
Prerequisites
Overview
The system rotating side 48V power supply does not require adjustment. The power supply can be checked for ripple and voltage level.
Procedure
- Ensure Axial drive is disabled via the service switch panel
prior to checking the power supply.note:
The power supply on this system does not require any adjustment. The power supply levels and ripple can be checked to confirm basic operation. The power supply and fuse board is located near the low channel end of DAS.
warning- The test points are located on the DAS backplane boards (both
low channel and high channel backplane board). Use below test points
to check the power supply level and ripple. If power supply does not
meet these specifications a replacement will be required.
Finalization
No finalization steps.