Collimator and Detector Heater PS Adjustment
Prerequisites
Overview
Procedure Effectivity:
1 Test Point Specifications
Procedure
- The detector heater power supply specifications are shown below:
2 Adjustment Procedure
Procedure
danger- Remove gantry side covers and top covers.
- Turn off Axial Drive and HVDC on the gantry’s Service Switch Panel (SSP).
- Rotate gantry until Collimator/Detector Power assembly reaches the 2 o’clock position.
danger- Remove four (4) nuts on Filament Power assembly safety cover and remove cover.
- Collimator P.S. output checks:
-
Connect positive voltmeter lead to +OUT terminal on power supply.
-
Connect negative voltmeter lead to -OUT terminal on power supply.
-
Turn voltage adjustment pot to adjust the output voltage to 24 ± 1 VDC.
-
- Reassemble the Gantry.
3 Finalization
No finalization steps.